page 8 Model 2002
2.3 Transducer Installation
The transducer may be installed in any orientation. Although the transducer is rugged and will
perform well in many harsh environments, the tube should be installed in a clean and careful
manner. The tube is configured with the vacuum fitting requested. If your vacuum environment
is highly contaminated or has unique fitting requirements, a Hastings filter or special adapter may
be needed. Please contact the Hastings Instruments Sales Department for assistance in your
system configuration.
2.4 Control Unit Installation
En
En
En
En
Environment:
vironment:
vironment:
vironment:
vironment:
Indoor use
Altitude up to 2000 meters
Operating temperature range from 5
0
to 40
0
C
Maximum relative humidity: 80% for temperatures up to 31
0
C decreasing linearly to 50%
relative humidity at 40
0
C
Installation category II
P
P
P
P
Panel Mount Instr
anel Mount Instr
anel Mount Instr
anel Mount Instr
anel Mount Instructions:
uctions:
uctions:
uctions:
uctions:
The control unit can be panel mounted. See detail on page 43. The hole dimensions on the panel
are 3.62" x 3.62"(92.8mm x 92.8 mm). Slide the neoprene gasket that was shipped with the
control unit onto the case from the back. Slide the controller through the panel cutout. Hold the
hardware against the side and tighten the two screws.
T
T
T
T
Transducer Cable
ransducer Cable
ransducer Cable
ransducer Cable
ransducer Cable Attachment:
Attachment:
Attachment:
Attachment:
Attachment:
The threaded connector attaches to the transducer. A finger tight connection is adequate for
proper operation. The 9-pin male “D” connector attaches to the back panel (See Figure 4.2).
The transducer cable connects to the left hand connector when looking at the back.
I/O Cable
I/O Cable
I/O Cable
I/O Cable
I/O Cable Attachment:
Attachment:
Attachment:
Attachment:
Attachment:
The mating male plug to the I/O connector is supplied with the unit. An I/O cable can be wired
with the 9-pin female “D” connector to use the analog output, digital alarms or the remote zero
functions. The connector will accept 20 gauge or smaller wire. The pinout is shown in Figure 4.2
(back panel figure). A detailed description of these pins is given in Section 4.
2.5 Initial Operation
Upon applying power to the control unit a pressure measurement will be given in Torr for nitrogen.
However, it is recommended that the user follow the instructions for zeroing and adjusting the
output at atmospheric pressure in Section 3.4 ("Cal Mode").
Содержание 2002
Страница 16: ...page 16 Model 2002 This page was intentionally left blank...
Страница 28: ...page 28 Model 2002 This page was intentionally left blank...
Страница 36: ...page 36 Model 2002 7 2 Relay Connector Pin Assignment 7 3 Operation...
Страница 42: ...page 42 Model 2002...
Страница 44: ...page 44 Model 2002 This page was intentionally left blank...
Страница 45: ...Model 2002 page 45 Diagrams and Drawings SECTION 11...
Страница 50: ...page 50 Model 2002...
Страница 51: ...Model 2002 page 51...
Страница 52: ...page 52 Model 2002...
Страница 53: ...Model 2002 page 53...
Страница 54: ...page 54 Model 2002...