![TELEDYNE API 465H Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/teledyne-api/465h/465h_user-manual_1080885018.webp)
16
Teledyne API Model 465H O
3
Monitor User Manual
06161H DCN7969
3.
For the
NEMA configuration
, the electrical connection must be hard-wired to the
3-pin connector on the mainboard labeled J23. AC power connection to the
monitor should be made with 12-14 AWG stranded copper wire, connected to the
monitor as follows:
a.
Earth Ground (green):
Connect the earth ground wire to the screw terminal lug on
the chassis as shown in the Figure below.
Do not
connect
the earth ground wire directly to the PCB connector on the
mainboard PCB.
b.
Line (Black):
Connect the Line wire to the connector on the mainboard
PCB labeled “Line”.
c.
Neutral (White):
Connect the Neutral wire to the connector on the
mainboard PCB labeled “Neutral”.
Figure 4-2. NEMA AC Power Connection to Monitor
4.
Connect the power cord to an appropriate power outlet (see the serial number tag
for correct voltage and frequency).
WARNING
Verify that the instrument is set up for proper line voltage and
frequency. Observe local electrical codes when connecting
power to the monitor.