AFA4000/2
Airflow Monitor / p.162
14.2
Contamination
When any part of the Auto Sash Control system is contaminated it must be either
removed and decontaminated or disposed of and replaced with a new part.
The local in-house Risk assessment, removal and disposal procedures (SOP) and COSHH
requirements must be observed whilst de-contaminating or disposing of the
contaminated part and should only be carried out by trained and competent parties.
Isolate the power to the controller before removing or replacing any contaminated parts
to avoid the risk of electric shock.
14.3
Maintenance
14.3.1
Faulty / replacement parts
If a faulty part has been identified (section 13.1), the Auto Sash Control system should be
isolated and marked as
Faulty
–
do not use
/
Out of service
until a replacement part has
been fitted. The fume cupboard can be operated manually during this period.
No special tools or parts are required for maintenance.
14.3.2
Validation
Once a replacement part has been fitted the Auto Sash Control system operation should
be fully tested to ensure safe operation.
A new tested label should be fitted or a new date should be applied to an existing label
to show that the system has been repaired and tested and is safe to use.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...