AFA4000/2
Airflow Monitor / p.152
13.
Troubleshooting
13.1
Fault finding
13.1.1
AFA4000/2
Problem
Check for
Screen blank
•
Check the power supply is securely plugged into the
monitor.
•
Check the power supply is securely plugged into the
mains power socket.
•
Check that there is 15 Vdc power supply on the monitor
terminals. If not, a wire could be broken or a cable sheath
caught.
Sensor Error message
•
Check the sensor cable is securely plugged into the
sensor.
•
Check the sensor cable is securely plugged into the
monitor.
•
Replace the sensor.
•
When the Sensor Error message is still displayed, replace
the sensor cable.
•
When the Sensor Error message is still displayed, replace
the monitor.
Audible alarm not
working
•
Check for screen messages. When
SETBACK
is displayed
the monitor has been remotely muted.
•
Check to see if the alarm has been disabled in the
(section 8.1.1.1).
RS485 Comms not
connected
•
•
Check the correct Protocol, Slave ID, Baud Rate and Parity
have been set.
•
Tx & Rx = 0. The AFA4000 is not connected to the
network.
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...