AFA4000/2
Airflow Monitor / p.112
Menu item
Instructions
Keyboard Click
Enable or disable the keyboard click.
Done
Return to the
Monitor config menu
(section 8.1.1).
8.1.1.2
Hours counter menu
Menu item
Instructions
Enable Counter
Enable or disable the hours counter.
Max Hours
Set the maximum hours for the hours counter.
(0
–
5000 hours)
.
Show Alarm
Select whether to display
the alarm, by selecting
Enable
or
not, by selecting
Disable
.
Reset Counter
Select
Yes
to reset the counter.
Done
Return to the
Monitor config menu
(section 8.1.1).
8.1.1.3
MODBUS settings menu
Menu item
Instructions
Slave ID
Set the Slave ID.
Baudrate
Set the Baudrate:
•
1200
•
2400
•
4800
•
9600
•
14400
•
19200
•
38400
•
57600
Parity
Set the parity:
•
None
•
Odd
•
Even
Done
Return to the
Monitor config menu
(section 8.1.1).
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...