AFA4000/2
Airflow Monitor / p.107
8.1.1
Monitor config menu
Note: A pop-up dialog box will show if a relay is already allocated to another option.
Menu item
Option
Instructions
Cal Config Menu
See
section 8.1.1.1.
Hours Counter
See section 8.1.1.2.
Input (1, 2 or 3)
Input activation
Select the activation type:
•
Closed contact
•
Open contact
•
None
•
Analog
Input function
Select the function:
•
Night Setback
•
Alarm Disable
•
None
•
Temperature
•
Fire Alarm
•
Mute
•
Fan Stop
•
Mains Fail
•
Sash Warning
•
High/Low
•
Sash High
•
Emergency
•
Ext. Alarm
•
Fire Alarm
Relay Output
(1, 2 or 3)
Relay activation
Select the activation type:
•
Closed contact
•
Open contact
Pushbutton (1,2,
or 3)
Pushbutton activation
Enable or disable the timer.
Pushbutton function
•
Fan
•
Ext Fan
•
D/F Fan
•
Sockets
•
S/B O/R
•
Purge
•
Light
•
UV Lights
•
Pump
•
Gas
•
High/Low
Speed
•
Emergency
Содержание AFA4000/2
Страница 1: ...AFA4000 2 Airflow monitor Installation and operating manual Issue 1 ...
Страница 50: ...AFA4000 2 Airflow Monitor p 50 Figure 21 AFA4000 2 typical connection diagram ...
Страница 51: ...AFA4000 2 Airflow Monitor p 51 Figure 22 AFA4000 2 AS typical connection diagram ...
Страница 52: ...AFA4000 2 Airflow Monitor p 52 5 4 Dimensions Figure 23 AFA4000 2 and AFA4000 2 AS dimensions ...
Страница 53: ...AFA4000 2 Airflow Monitor p 53 Figure 24 SM7 airflow sensor dimensions ...
Страница 64: ...AFA4000 2 Airflow Monitor p 64 Figure 34 Side views of a rack and pinion drive typical installation ...
Страница 87: ...AFA4000 2 Airflow Monitor p 87 Figure 61 Adjusting the personnel sensor beam angle away from the fume hood ...
Страница 90: ...AFA4000 2 Airflow Monitor p 90 6 10 Auto Sash System Control Box general wiring diagram ...
Страница 127: ...AFA4000 2 Airflow Monitor p 127 ...
Страница 140: ...AFA4000 2 Airflow Monitor p 140 Figure 69 Typical connection diagram ...
Страница 144: ...AFA4000 2 Airflow Monitor p 144 Figure 70 Connection diagram with RS232 485 converter ...