8.  If the sample is accepted, lower the sash by half and 

measure the face velocity using a calibrated instrument. 
Using the + / - buttons enter the measured face velocity 
and press Enter, the monitor will then sample the airflow 
for 5 seconds.

9.  If the calibration is successful the controller will return the 

Main Menu, select RUN to go to normal operating mode 
and check the velocity reading is accurate and stable.

10.  If the airflow sample is unstable the controller will 

display “Deviations too High”, follow the instructions to 
repeat the sample or quit the calibration.

11.  The controller will display “Increase higher airflow 

sample” if the second sample value is too close to the 
first value entered, close the sash a little and repeat 
the higher sample. The minimum difference between 
the samples that the controller will accept is 0.3m/
sec (60fpm).

12.  The controller will display “Sensor diff too low” if the 

controller doesn’t detect any difference in the sensor 
output between the 2 airflow samples, check that the 
sensor hose is connected and repeat the calibration.

13.  If the calibration is stable and accurate Press and 

Hold the Enter button for 5 seconds until the Main 
Menu is displayed.

14.  Using the + / - buttons select SET UP, then select 

CONFIGURE, then enter the password (the factory 
default password is 0-0-0-0), press Enter to continue.

15.  Using the + / - buttons select ECON CONFIGURE, then 

select MANUAL / AUTO and press Enter, then select 
AUTO and press Enter to continue, then select DONE 
and press Enter.

16.  The controller will return to the MAIN MENU, select 

RUN to go to normal operating mode.

Calibration Tips

1.  Ensure that the sensor hose is connected correctly at the 

side wall and rear of the sensor.

2.  Ensure that the extract fan is running and the Fume 

Cupboard is balanced before calibrating the monitor.

3.  Take extra time (at least 15 to 20 seconds) to wait for the 

airflow to settle before capturing the airflow samples. 
This will minimize the chance of a calibration error due 
to turbulence or fluctuations.

4.  Avoid movement in front of the Fume Cupboard whilst 

calibrating the monitor.

5.  The Low and High air samples must be at least 0.3m/sec 

(60 fpm) apart to calibrate the monitor. This is to avoid 
inaccuracy in the calibration sue to insufficient difference 
between the samples. The minimum difference can be 
changed in the “Low High diff” parameter in the  
Cal Config menu.

6.  Do not use fully open and fully closed sash positions for 

the calibration points, the recommended positions are 
normal operating height (e.g. 500mm 20”) for the first 
sample and approximately half the sash opening (e.g. 
250mm 10”) for the second sample.