background image

 

14

4.0 SM6 Airflow Sensor Installation  

 
 
 

 

It is very important to position the SM6 airflow sensor in the correct position to give  long term stable reading of the 
face velocity. Please read the  INSTALLATION NOTES below and if in doubt contact us for further advice. 
 

INSTALLATION  NOTES :- 
 

1.  The SM6 sensor must be positioned where it can " see " the room pressure of the laboratory.The back connection    
     spigot of the sensor is designed to accept a 25mm OD tube which should be connected to the inner chamber of  
     the fume cupboard. (This tube and fittings is known as the " vent kit " ) 
     The 

ideal position

 for the end of the 25mm tube for most fume cupboards is 100mm back from the sash glass  

     and 100mm higher than the normal sash opening height through the inner side wall. 
 
2.  If possible mount the sensor on the front of the fume cupboard and use a short length of tube. Tube lengths of  
     more than 1metre or smaller diameter will restrict the airflow through the sensor. This will lead to too much  
     sensitivity being required to  calibrate the unit which can lead to some instability of the reading or incorrect  
     readings at low velocitiesl. 
 
3.  For fume cupboards with a single skin side wall or a double skin with a small gap between them it may not be 
     possible to achieve the ideal sensing position using a flexible tube. 
     With a single skin side wall it is possible to fix the sensor on the outside of the fume cupboard and connect directly  
     to the inner chamber in the ideal position. This method can only be used for up to two fume cupboards when they  
     are positioned side by side ( using the two outer walls ). 
     An alternative method is to mount the sensor on the front of the fume cupboard and connect using a short flexible  
     tube to a rigid wall tube attached to the inner side wall. The open end of this rigid wall tube should be positioned in  
     the ‘ideal position’  ie. 100mm back from the sash and 100mm higher than the normal sash opening. 
     Fume cupboards with a high internal height can present a difficulty because the tube length to reach the ideal  
     position may be longer than 1 metre. In this case it is better to use a tube no longer than 1 metre which may result  
     in a sensing position higher than the ideal. 
     When fitting a sensor to a ’narrow wall’ fume cupboard for the first time it may be necessary to try various  
     combinations of rigid and flexible tube to find the best combination and position. 
      
4.  The sensor should not be mounted in a position were it is subject to draughts from the laboratory air input or  
     ventilation system.

 

 
 

 

 

100mm 

100mm 

Norm al Sash  

working height. 

eg. 500mm 

SM6 Sensor 
Open to the 
laboratory 

Fume cupboard 

outer skin 

Fume cupboard 

inner skin 

25mm Tube 

25mm 

Female adapter 

25mm

Male 

bush 

SM6 Airflow Sensor 

( See exploded view below ) 

Содержание AFA1000 / 1 - Mk2

Страница 1: ...Digital display 3 Relay inputs 3 Relay outputs Com port Used for alarm indication and monitoring on Fume Cupboards Model AFA1000 2 Mk2 Digital display 3 Relay inputs 3 Relay outputs 3 Pushbuttons Com...

Страница 2: ...Passwords in the Main Menu or alternatively use a Laptop connected to the Com port and the Upload Download software provided Velocity Bar Graph or Alarm Time Line Velocity display m sec or fpm Control...

Страница 3: ...R2 R3 SENSOR SOCKET 0 10V ECON OUTPUT 15V DC POWER SUPPLY INPUT 0 10V AUX OUTPUT s w Sash Proximity switch OUTPUT RELAYS COMMS PORT RELAY INPUTS INPUT 1 INPUT 2 INPUT 3 Bn Bk Bl Sash Micro switch REF...

Страница 4: ...representing events that have occurred during the past hour The alternative to the event time line is a dynamic bar graph representing the face velocity The display shows a Horn icon with line through...

Страница 5: ...the calibration and configuration menu or to select options or values External Connections the alarm unit will have the following connection points Input 1 volt free relay input configurable for norma...

Страница 6: ...a Fan stop start button for the fume cupboard extract fan or as Night Set back Override Pushbutton 2 This button can be configured to act as a Lights ON OFF or UV Lights On OFF for the fume cupboard o...

Страница 7: ...e configuration of the alarm NOTE If you enter the Calibration or Configure Menu by accident press the buttons at the same time to escape back to the Main Menu The menus and sub menus are in plain lan...

Страница 8: ...ime the Low air alarm resets automatically High airflow If configured High Air toggles on off with display Audible alarm sounds can be muted via Enter pushbutton Audible alarm Mute When the audible al...

Страница 9: ...ctivated Night Set Back Icon is displayed Red LED on Flashing Reduced Low air alarm if configured Audible alarm muted Mute Icon shown on display External alarm When input configured as External alarm...

Страница 10: ...isables Night setback operation in Override On condition Note In the On condition the unit ignores any Night Set back input signals and remains in this condition until Off is selected Pushbutton 2 Pum...

Страница 11: ...setup press ENTER to continue and in the Main Menu use the buttons on the alarm facia select SETUP and then press the ENTER button 7 In the Setup Menu select CALIBRATION and press the ENTER button 8...

Страница 12: ...individual lowest point on the grid For most fume cupboards this low point is on the bottom row in the centre and is a convenient position to measure and for future reference when checking the alarm...

Страница 13: ...ensor Hole for Vent Tube SM6 Sensor Dimensions Rear view SM6 Sensor Panel Cutout Dimensions Front view NOT TO SCALE 2 x Holes for self tapping fixing screw 4mm Fixing Hole provided on SM6 sensor Panel...

Страница 14: ...ng position using a flexible tube With a single skin side wall it is possible to fix the sensor on the outside of the fume cupboard and connect directly to the inner chamber in the ideal position This...

Страница 15: ...SM6 Airflow Sensor R1 R2 R3 SENSOR SOCKET 0 10V ECON OUTPUT 15V DC POWER SUPPLY INPUT 0 10V AUX OUTPUT s w Sash Proximity switch OUTPUT RELAYS COMMS PORT RELAY INPUTS INPUT 1 INPUT 2 INPUT 3 Bn Bk Bl...

Страница 16: ...oximity switch Optional SM6 Airflow Sensor R1 R2 R3 SENSOR SOCKET 0 10V ECON OUTPUT 15V DC POWER SUPPLY INPUT 0 10V AUX OUTPUT s w Sash Proximity switch OUTPUT RELAYS COMMS PORT RELAY INPUTS INPUT 1 I...

Страница 17: ...F all other warranties and is subject to the conditions and LIMITATIONS stated herein NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE THE EXCLUSIVE RE...

Отзывы: