2.1  Quick Start Installation  

 

Follow the instructions below for installing and commissioning the unit. :- 
 

1.  Fit the alarm to the Fume Cupboard using the cut-out details provided with the unit 

 

  --- see page 13 

 

 

2.  Fit the airflow sensor to the Fume Cupboard using the cut out and installation details 

provided --- see page 13  & 14 

 

3.  Connect the ‘telephone style’  airflow sensor plug-in cable to the sensor and the back 

of the alarm unit --- see typical connection diagram on page 15 & 16 

 

4. 

Plug in the power adapter to a Mains AC power socket and connect the flying lead to 
the alarm unit  --- see typical connection diagram on page 15 & 16

 

 

5. 

Power up the unit and wait at least 30 secs while the sensor temperature stabilises.

 

 

6. 

If the unit has not been calibrated the unit will display ‘ Requires setup ’ – press 
ENTER to continue and in the Main Menu use the +/- buttons on the alarm facia 
select ‘ SETUP ‘ and then press the ENTER button.

 

 

7. 

In the Setup Menu  select ‘CALIBRATION’ and press the ENTER button

 

 

8. 

At this stage you will be requested to enter the PASSWORD. Use the +/- buttons to 
select the individual digits in turn and then press ENTER. 

 

If the password is correct the unit will go to the calibration mode. If the password is 
not correct you will be requested to try again --- on the third wrong password entry the 
calibration menu will lock out for 10 mins

 

 

9. 

When in the calibration mode follow the instructions on the display screen to carry out 
the calibration of the unit. See ‘Calibration Notes’ below for hints on successful 
calibration.

 

       When the calibration is complete the unit will return to the Main Menu.

 

 

10. 

Use the +/- buttons on the alarm facia select ‘ RUN ’ and then press the ENTER 
button.

 

 
 
The unit will now function and display the measured Fume Cupboard face 
velocity 

 
 
 
 
 
 
 
 
 
 
 
 

Содержание AFA1000 / 1 - Mk2

Страница 1: ...Digital display 3 Relay inputs 3 Relay outputs Com port Used for alarm indication and monitoring on Fume Cupboards Model AFA1000 2 Mk2 Digital display 3 Relay inputs 3 Relay outputs 3 Pushbuttons Com...

Страница 2: ...Passwords in the Main Menu or alternatively use a Laptop connected to the Com port and the Upload Download software provided Velocity Bar Graph or Alarm Time Line Velocity display m sec or fpm Control...

Страница 3: ...R2 R3 SENSOR SOCKET 0 10V ECON OUTPUT 15V DC POWER SUPPLY INPUT 0 10V AUX OUTPUT s w Sash Proximity switch OUTPUT RELAYS COMMS PORT RELAY INPUTS INPUT 1 INPUT 2 INPUT 3 Bn Bk Bl Sash Micro switch REF...

Страница 4: ...representing events that have occurred during the past hour The alternative to the event time line is a dynamic bar graph representing the face velocity The display shows a Horn icon with line through...

Страница 5: ...the calibration and configuration menu or to select options or values External Connections the alarm unit will have the following connection points Input 1 volt free relay input configurable for norma...

Страница 6: ...a Fan stop start button for the fume cupboard extract fan or as Night Set back Override Pushbutton 2 This button can be configured to act as a Lights ON OFF or UV Lights On OFF for the fume cupboard o...

Страница 7: ...e configuration of the alarm NOTE If you enter the Calibration or Configure Menu by accident press the buttons at the same time to escape back to the Main Menu The menus and sub menus are in plain lan...

Страница 8: ...ime the Low air alarm resets automatically High airflow If configured High Air toggles on off with display Audible alarm sounds can be muted via Enter pushbutton Audible alarm Mute When the audible al...

Страница 9: ...ctivated Night Set Back Icon is displayed Red LED on Flashing Reduced Low air alarm if configured Audible alarm muted Mute Icon shown on display External alarm When input configured as External alarm...

Страница 10: ...isables Night setback operation in Override On condition Note In the On condition the unit ignores any Night Set back input signals and remains in this condition until Off is selected Pushbutton 2 Pum...

Страница 11: ...setup press ENTER to continue and in the Main Menu use the buttons on the alarm facia select SETUP and then press the ENTER button 7 In the Setup Menu select CALIBRATION and press the ENTER button 8...

Страница 12: ...individual lowest point on the grid For most fume cupboards this low point is on the bottom row in the centre and is a convenient position to measure and for future reference when checking the alarm...

Страница 13: ...ensor Hole for Vent Tube SM6 Sensor Dimensions Rear view SM6 Sensor Panel Cutout Dimensions Front view NOT TO SCALE 2 x Holes for self tapping fixing screw 4mm Fixing Hole provided on SM6 sensor Panel...

Страница 14: ...ng position using a flexible tube With a single skin side wall it is possible to fix the sensor on the outside of the fume cupboard and connect directly to the inner chamber in the ideal position This...

Страница 15: ...SM6 Airflow Sensor R1 R2 R3 SENSOR SOCKET 0 10V ECON OUTPUT 15V DC POWER SUPPLY INPUT 0 10V AUX OUTPUT s w Sash Proximity switch OUTPUT RELAYS COMMS PORT RELAY INPUTS INPUT 1 INPUT 2 INPUT 3 Bn Bk Bl...

Страница 16: ...oximity switch Optional SM6 Airflow Sensor R1 R2 R3 SENSOR SOCKET 0 10V ECON OUTPUT 15V DC POWER SUPPLY INPUT 0 10V AUX OUTPUT s w Sash Proximity switch OUTPUT RELAYS COMMS PORT RELAY INPUTS INPUT 1 I...

Страница 17: ...F all other warranties and is subject to the conditions and LIMITATIONS stated herein NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE THE EXCLUSIVE RE...

Отзывы: