Operating Procedure
1. Slide open the battery door at the back of the unit.
2. Install the 4 1.5V AAA battery and replace the battery door.
3. Plug the sensor into the connector at the top of the unit.
4. Turn on the TeslaMeter 2000 by pressing the POWER switch for at least 1 sec.
The TeslaMeter 2000 is manually operated by 5 switches.
Power
On and Off switch.
Press once for at least 1 sec: instrument turns On.
Press again:
instrument turns Off.
Note: After turning the sensor off, wait at least 5 seconds before turning it on again.
If the “POWER” switch of the TeslaMeter 2000 is not being used or if not connected to USB, the
TeslaMeter 2000 switches off automatically after 10 minutes. The display shows “OFF” for
about 1sec. to indicate that the instrument is switched off. When the battery is too low, the
display indicates “LOBAT” and a few seconds later, the TeslaMeter 2000 switches Off.
When no probe sensor is attached to the connector at the top of the unit, the TeslaMeter 2000
switches Off a few seconds after being switched On.
Range
The values of measurement are displayed in three ranges:
1.
19.999 mT
2.
199.99 mT
3.
1999.9 mT
When the TeslaMeter 2000 is switched On, it is set on the range last used.
By pressing “RANGE” the unit toggles from the last range used to Auto ranging (display shows
blinking dot) to 1999.9 (display shows 0.0) to 199.99 (display show 0.00) to 19.999 (display
shows 0.000).
Note:
If the magnetic field being measured is greater than the chosen range, then the LED will
read FUL.
Zero Cal
This button allows one to calibrate the Hall sensor offset. The user has to place the probe in the
Zero Gauss Chamber, then press the button “ZERO CAL”. The Hall sensor offset is then
measured and memorized. The new set of Offsets remain after switching off the TelsaMeter