Release notes
Incompatible operation
When applying 3 Gb/s signals to any instrument that does not support the
signal formats, the unit indicates that the signal input is not standard. Applying
unsupported video signals may result in a frozen PICT display.
SNMP messages
Commands returning the audCurOutput OID return the string “Embed 7 & 8” for
embedded audio channel pairs 9 & 10, 11 & 12, 13 & 14, and 15 & 16.
Sweep
The waveform style must be set to Overlay to get 2 line or 2
fi
eld sweep.
USB
Always press the MAIN button and select USB Status to safely “Unmount”
and remove the USB memory device from the USB port. You risk permanent
loss of any
fi
les saved on the USB device if you do not use the “Unmount”
feature.
If there are more than 32
fi
les saved in a Capture or Presets directory on the
USB device, you can only view the
fi
rst 32
fi
les.
Waveform mode cursor for
XYZ color space
The voltage cursors are given in millivolt scale, but the XYZ color space is
referenced in hex scale. To see the values for 000 to FFF, change the graticule
settings to max and percent.
Web browser
The Web browser applet will not resize when running in some versions of
the Netscape browser and in some non-Windows Operating Systems (for
example, Solaris).
We recommend using Java Runtime Engine (JRE) version 1.6 or above.
WFM7200 Waveform Monitor v3.0.X Release Notes
5