Calibration — Type Z
COM PARISON VO LTAG E Helidial. At null, the Helidial
dial reading should not be further than '/2 minor division
(which is 5 millivolts) from the 10.000 position.
Disconnect the voltmeter, remove the plug-in extension
and insert the Z Unit directly into the oscilloscope plug-in
compartment. Install the left side panel on the oscilloscope.
®
5-7