background image

Preface

This document helps customers with data security concerns to sanitize or remove
memory devices from TLA7S08 and TLA7S16 Serial Analyzer Modules.

These products have data storage (memory) devices and data output devices (USB
ports). These instructions tell how to do the following:

Clear or sanitize the memory devices

Clear or sanitize an instrument that is not functioning

Products

The following Tektronix products are covered by this document:

TLA7S08

TLA7S16

Related Documents

The

TLA7S08 and TLA7S16 Serial Analyzer Modules Service Manual

,

part number 071-2888-XX, is available on the Tektronix Web site at
www.tektronix.com/manuals.

Terms

The following terms may be used in this document:

Clear.

This removes data on media/memory before reusing it in a secured area.

All reusable memory is cleared to deny access to previously stored information
by standard means of access.

Erase.

This is equivalent to clear.

Media storage/data export.

Various devices that are used to store or export data

from the instrument, such as a USB port.

Nonvolatile memory.

Data is retained when the instrument is powered off.

Remove.

This is a physical means to clear the data by removing the memory

device from the instrument. Instructions are available in the product service
manual.

Sanitize.

This eradicates the data from media/memory so that the data cannot be

recovered by other means or technology. This is typically used when the device
will be moved (temporarily or permanently) from a secured area to a non-secured
area.

Scrub.

This is equivalent to sanitize.

User-modi

fi

able.

The memory device can be written to by the user during normal

instrument operation, using the instrument’s user interface or remote control.

Volatile memory.

Data is lost when the instrument is powered off.

TLA7S08 and TLA7S16 Declassi

fi

cation and Security Instructions

iii

Содержание TLA7S08

Страница 1: ...TLA7S08 and TLA7S16 Serial Analyzer Modules Declassification and Security Instructions www tektronix com 071 2189 00...

Страница 2: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Страница 3: ...ents Preface iii Clear and Sanitize Procedures 1 Memory Devices 1 Data Export Devices 2 Troubleshooting 3 How to Clear a Non Functional Instrument 3 TLA7S08 and TLA7S16 Declassification and Security I...

Страница 4: ...Table of Contents ii TLA7S08 and TLA7S16 Declassification and Security Instructions...

Страница 5: ...ess to previously stored information by standard means of access Erase This is equivalent to clear Media storage data export Various devices that are used to store or export data from the instrument s...

Страница 6: ...Preface iv TLA7S08 and TLA7S16 Declassification and Security Instructions...

Страница 7: ...tize FLASH 2 M X 16 Stores instrument firmware and FPGA data No Programmed at the factory no user data LPU board Remove LPU board and either securely store it or destroy it NVRAM 128 K X 8 Stores inst...

Страница 8: ...ce Clock cable from the output port on the front panel See below VXI backplane Interfaces with TLA mainframe No Instrument function and application software Backplane connectors Power off the TLA main...

Страница 9: ...ar a Non Functional Instrument If your instrument is not functioning and you need to clear it remove the power source from the instrument for at least 20 seconds TLA7S08 and TLA7S16 Declassification a...

Страница 10: ...Troubleshooting 4 TLA7S08 and TLA7S16 Declassification and Security Instructions...

Отзывы: