Adjustment Procedures
5- 2
TLA7AXX/TLA7NAX Logic Analyzer Module Service Manual
To perform the self calibration procedure, do the following:
1.
Ensure that the instrument has had a 30-minute warm up before attempting
the self calibration, and that the logic analyzer application is running.
2.
Disconnect any probes connected to the logic analyzer module.
3.
Select Calibration and Diagnostics from the System menu.
4.
Select the Self Calibration tab page.
5.
Select the logic analyzer module.
6.
Click the Run button to start the self calibration.
The self calibration takes several minutes to complete, depending on the
number of channels in the module. Upon successfully completing the self
calibration, the module status changes from Running to Calibrated, and the
Date and Time field is set to the present.
Содержание TLA7AA Series
Страница 5: ......
Страница 11: ...Table of Contents vi TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Страница 15: ...Service Safety Summary x TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Страница 21: ...Preface xvi TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Страница 25: ...Introduction xx TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Страница 27: ...Specifications 1 2 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Страница 43: ...Operating Information 2 16 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Страница 49: ...Theory of Operation 3 6 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Страница 53: ...Performance Verification 4 4 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Страница 95: ...Troubleshooting 6 40 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Страница 101: ...Options 7 4 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Страница 103: ...Electrical Parts List 8 2 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Страница 107: ...Diagrams 9 4 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...