Appendix D: Performance Verification
THS 710 & THS 720 User Manual
D–15
3. Set the leveled sine wave generator frequency to 100 MHz.
4. Set the TekScope instrument VOLTS/DIV to 500 mV/div.
5. Set the TekScope instrument SEC/DIV to 10 ns/div.
6. Set the leveled sine wave generator output level to approximately
50 mV
p-p
so that the measured amplitude is approximately
500 mV. (The measured amplitude can fluctuate around 500 mV.)
7. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as
necessary and then check that triggering is stable.
8. Change the TekScope instrument setup using the following steps:
SCOPE
TRIGGER
Slope
\ (falling edge)
9. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as
necessary and then check that triggering is stable.
10. Change the TekScope instrument setup using the following steps:
SCOPE
TRIGGER
Slope
/ (rising edge)
11. Proceed to the next test to check the channel 2 edge trigger
sensitivity.
Содержание TekScope THS 710
Страница 12: ...Getting Started...
Страница 13: ......
Страница 24: ...Operating Basics...
Страница 25: ......
Страница 65: ...Application Examples 2 40 THS 710 THS 720 User Manual...
Страница 66: ...Reference...
Страница 67: ......
Страница 69: ...Introduction to Reference 3 2 THS 710 THS 720 User Manual...
Страница 131: ...VERTICAL Controls 3 64 THS 710 THS 720 User Manual...
Страница 132: ...Appendices...
Страница 133: ......
Страница 157: ...Appendix C Accessories C 6 THS 710 THS 720 User Manual...
Страница 179: ...Appendix D Performance Verification D 22 THS 710 THS 720 User Manual...
Страница 181: ...Appendix E General Care and Cleaning E 2 THS 710 THS 720 User Manual...
Страница 182: ...Glossary and Index...
Страница 183: ......