Appendix D: Performance Verification
THS 710 & THS 720 User Manual
D–11
3. Set the TekScope instrument VOLTS/DIV to 100 mV/div.
4. Set the TekScope instrument SEC/DIV to 10
s/div.
5. Set the leveled sine wave generator frequency to 50 kHz.
6. Set the leveled sine wave generator output level so the peak-to-
peak measurement is between 599 mV and 601 mV.
7. Set the leveled sine wave generator frequency to 60 MHz if you
are checking a THS 710 or to 100 MHz if you are checking a
THS 720.
8. Set the TekScope instrument SEC/DIV to 10 ns/div.
9. Check that the peak-to-peak measurement is
≥
425 mV.