Performance Veri
fi
cation
10.
Disconnect the test setup.
11.
Repeat steps 1 through 10 for all input channels.
Check Sample Rate
Accuracy and Delay Time
Accuracy
This test checks the time base accuracy.
1.
Set up the oscilloscope using the following table:
Push menu button
Select menu option
Select setting
Default Setup
—
—
Channel
1
Probe
1X
2.
Connect the oscilloscope to the time mark generator as shown in the following
fi
gure:
3.
Set the time mark generator period to
10 ms
.
4.
Set the oscilloscope
Vertical Scale
(volts/division) to
500 mV/div
.
5.
Set the oscilloscope Main
Horizontal Scale
(seconds/division) to
1 ms/div
.
6.
Push
Set To 50%
.
7.
Use the
Vertical Position
control to center the test signal on screen.
8.
Use the
Horizontal Position
control to set the position to
10.00 ms
.
9.
Set the oscilloscope
Horizontal Scale
(seconds/division) to
250 ns/div
.
32
TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual
Содержание TDS1001C-EDU
Страница 12: ...Preface x TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Страница 30: ...Specifications 18 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Страница 32: ...Where to Find Operating Information 20 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Страница 52: ...Performance Verification 40 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Страница 60: ...Adjustment Procedures 48 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Страница 85: ...Maintenance TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual 73...
Страница 96: ...Maintenance 84 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...
Страница 110: ...Replaceable Parts 98 TDS2000C and TDS1000C EDU Series Oscilloscope Service Manual...