Performance Tests
TDS 820 Performance Verification
1Ć37
First, set the vertical and horizontal
scale and the generator
frequency and amplitude.
Second, measure the
waveform period.
Figure 1-15: Measurement of Accuracy — Single Channel Delta Time Measurement at 1 ns
2.
Confirm that time measurements are within limits for accuracies at 1 ns:
a.
Display the test signal:
H
Press SHIFT; then ACQUIRE MENU. Now press the main-menu
button Mode; then press the side-menu button Average.
b.
Check single channel delta-time measurement accuracy at 1 ns
against limits:
H
Press MEASURE.
H
Press the main-menu button High-Low Setup; then press the
side-menu button Min-Max.
H
Press the main-menu button Select Measurement for Chx.
H
Press the side-menu button –more–, until PERIOD appears in
the side menu. Press PERIOD.
H
Press HORIZONTAL MENU.
H
Press the main-menu button Time Base Position; then adjust
the general purpose knob clockwise to its minimum setting
(–1.5 ns for the standard instrument and 16 ns for Option 1D).
H
Press CLEAR MENU.
H
CHECK that the period readout is within 984 ps to 1.016 ns (see
Figure 1-15).
Содержание TDS 820
Страница 4: ......
Страница 6: ...Preface ii...
Страница 12: ...Safety viii...
Страница 13: ...Performance Verification Procedures...
Страница 14: ......
Страница 70: ...Performance Tests Performance Verification 1 56...
Страница 71: ...Performance Tests TDS 820 Performance Verification 1 57...
Страница 72: ...Specifications Replace this page with the tab divider of the same name...
Страница 73: ......
Страница 74: ...TDS 820 Performance Verification 2 1 Appendix B Specification...
Страница 80: ...Product Description Appendix B Specification TDS 820 Performance Verification 2 7...
Страница 81: ...Product Description Appendix B Specification Specification 2 8...
Страница 91: ...Warranted Characteristics Appendix B Specification Specification 2 18...
Страница 94: ...Typical Characteristics Appendix B Specification TDS 820 Performance Verification 2 21...