TDS 420A, TDS 430A & TDS 460A Service Manual
4- 15
Performance Tests
This section contains procedures for checking that the TDS 400A Digitizing
Oscilloscopes perform as warranted.
The procedures are arranged in four logical groupings:
Signal Acquisition System
Check
s
, Time Base System Checks, Triggering System Checks
, and O
utput Ports
Checks
. They check all the characteristics that are designated as checked in
Chapter 1,
Specifications.
(The characteristics that are checked appear in
boldface
type under
Warranted Characteristics
in Chapter 1.) You can use the
form at the end of this section as a test record.
The procedures are arranged in four logical groupings:
Signal Acquisition System
Check
s
, Time Base System Checks, Triggering System Checks
, and O
utput Ports
Checks
. They check all the characteristics that are designated as checked in
Chapter 1,
Specifications.
(The characteristics that are checked appear in
boldface
type under
Warranted Characteristics
in Chapter 1.) You can use the
form at the end of this section as a test record.
Read
Performance Verification Procedures
that start on page NO TAG. Also, if
you are not familiar with operating this digitizing oscilloscope, read
Operating
Information
in Chapter 2 of the service manual or read the user manual before
doing any of these procedures.
STOP
. These procedures extend the confidence level provided by the basic
procedures described on page 4--5. The basic procedures should be done first,
and then these procedures performed if desired.
Prerequisites
The tests in this chapter comprise an extensive, valid confirmation of perfor-
mance and functionality when the following requirements are met:
H
The cabinet must be installed on the digitizing oscilloscope.
H
You must have performed and passed the procedures under
Self Tests,
found
on page 4--5, and those under
Functional Tests,
found on page 4--7.
H
A signal-path compensation
must
have been done within the recommended
calibration interval and at a temperature within
±
5
_
C of the present
operating temperature. (If at the time you did the prerequisite
Self Tests,
the
temperature was within the limits just stated, consider this prerequisite met.)
Содержание TDS 420A
Страница 16: ...Service Safety Summary xii TDS 420A TDS 430A TDS 460A Service Manual ...
Страница 36: ...Nominal Traits 1 14 TDS 420A TDS 430A TDS 460A Service Manual ...
Страница 44: ...Warranted Characteristics 1 22 TDS 420A TDS 430A TDS 460A Service Manual ...
Страница 54: ...Installation 2 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Страница 64: ...Operating Information 2 16 TDS 420A TDS 430A TDS 460A Service Manual ...
Страница 72: ...Performance Verification Procedures 4 4 TDS 420A TDS 430A TDS 460A Service Manual ...
Страница 160: ...Maintenance 6 8 TDS 420A TDS 430A TDS 460A Service Manual ...
Страница 226: ...Removal and Installation Procedures 6 74 TDS 420A TDS 430A TDS 460A Service Manual ...
Страница 243: ...Troubleshooting TDS 420A TDS 430A TDS 460A Service Manual 6 91 J2 Pin 1 Figure 6 42 A02 Display Connector J2 ...
Страница 260: ...Options and Accessories 7 6 TDS 420A TDS 430A TDS 460A Service Manual ...
Страница 262: ...Electrical Parts List 8 2 TDS 420A TDS 430A TDS 460A Service Manual ...
Страница 272: ...Diagrams 9 10 TDS 420A TDS 430A TDS 460A Service Manual ...
Страница 287: ...TDS 420A TDS 430A TDS 460A Service Manual 10 15 1 2 3 4 5 6 7 3 8 9 10 11 Figure 10 5 Cables and routing B080000 Up ...