background image

Reference Notes

Measuring Differential Current

You can place two conductors in a current probe to provide differential or null
current measurement. (See Figure 11.) This avoids the necessity of using two
current measurement systems with a differential oscilloscope.

Figure 11: Measuring two conductors

WARNING.

To avoid injury or loss of life from shock or

fi

re, do not put more than

one uninsulated conductor at a time in the probes.

An insulated conductor is de

fi

ned as any conductor that is surrounded by an

insulating material that is capable of isolating the voltage present on the conductor.
Note that lacquer coatings like those typically found on transformer windings do
not provide suf

fi

cient, reliable insulation for use with current probes. The lacquer

coating can be easily nicked or damaged, compromising the insulating capabilities
of the lacquer coating.

Do not force the slide closed. Damage to the probe may result. If you cannot close
the slide around the conductor(s), either reduce the number of conductors you are
measuring, or, if possible, take your measurement on a smaller conductor.

1.

Orient the two conductors under test so that the polarities oppose each other.

2.

Clamp the current probe around the two conductors as shown. Be careful not
to pinch a conductor in the probe jaws.

24

TCPA300/400 Ampli

fi

ers and TCP300A/400 Series Current Probes User Manual

Содержание TCP300A

Страница 1: ...xx TCPA300 400 Amplifiers TCP300A 400 Series AC DC Current Probes ZZZ User Manual P077118301 077 1183 01...

Страница 2: ......

Страница 3: ...TCPA300 400 Amplifiers TCP300A 400 Series AC DC Current Probes ZZZ User Manual xx This document applies for firmware version 1 0 and above www tektronix com 077 1183 01...

Страница 4: ...in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc Additional tra...

Страница 5: ...taxes and any other charges for products returned to any other locations This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and...

Страница 6: ......

Страница 7: ...Probe Slide 10 Degaussing and Autobalancing the Current Probe 12 DC Measurements 13 AC Measurements 15 Control Summary 17 TCPA300 and TCPA400 Controls 18 Reference Notes 23 Degaussing a Probe with an...

Страница 8: ...se of an Error Code 47 Shutdown Error 47 Specifications 49 Warranted Specifications 49 Nominal and Typical Characteristics 50 Mechanical Characteristics 51 Environmental Characteristics 53 Performance...

Страница 9: ...18 Increasing probe sensitivity 34 Figure 19 Setup for measuring charging current 36 Figure 20 Charge current waveforms 36 Figure 21 Measuring inductance with a low impedance source 37 Figure 22 Line...

Страница 10: ...Troubleshooting 42 Table 7 Amplifier error codes 46 Table 8 Warranted TCPA300 and TCPA400 specifications 49 Table 9 Nominal and typical amplifier characteristics 50 Table 10 TCPA300 and TCPA400 mecha...

Страница 11: ...ct the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement instrument Ground the product This product is grounded through the groun...

Страница 12: ...ices that could result in injury or loss of life CAUTION Caution statements identify conditions or practices that could result in damage to this product or other property Symbols and terms on the prod...

Страница 13: ...transient burst immunity IEC 61000 4 5 2001 Power line surge immunity IEC 61000 4 6 2003 Conducted RF immunity IEC 61000 4 11 2004 Voltage dips and interruptions immunity 4 EN 61000 3 2 2006 AC power...

Страница 14: ...10 1 Safety requirements for electrical equipment for measurement control and laboratory use UL61010B 2 032 Particular requirements for hand held current clamps for electrical measurement and test Can...

Страница 15: ...condensation occurs only when the product is out of service Pollution Degree 3 Conductive pollution or dry nonconductive pollution that becomes conductive due to condensation These are sheltered locat...

Страница 16: ...encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately This symbol indicates that this product complies with t...

Страница 17: ...tion The Glossary and Index are provided as quick reference locators for important information Service Manual The Performance Verification and Adjustment procedures are located in the Service Manual a...

Страница 18: ...Preface xii TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...

Страница 19: ...sociated probes provide these features Simultaneous DC and AC current measurements up to 750 A peak High sensitivity One button autobalancing and probe degaussing No adjustments needed to match a curr...

Страница 20: ...TCPA400 Amplifier TCP404XL 750 A 1 2 MHz 1 500 A continuous 750 A DC derated with duty cycle Oscilloscope An oscilloscope displays the output from the current measuring system A 50 cable is included t...

Страница 21: ...ctory on delivery C3 Provides factory calibration certification on delivery plus two more years of calibration coverage Throughout the coverage period the instrument will be calibrated according to it...

Страница 22: ...er the following optional accessories for the amplifiers and probes One turn 50 HF current loop Two versions are available one for each style of probe The current loops are used in the performance ver...

Страница 23: ...he probe cover to hold your probe in a convenient place at your bench or workstation when you are not using it You can attach the probe cover to the side of the bench to keep the probe off of your wor...

Страница 24: ...s have a grounded shield between the conductor under test and the current sense transformer Any capacitively coupled current will then flow in the ground instead of the transformer windings When you t...

Страница 25: ...ent Probes one of each size For example you can store a TCP305A and a TCP303 probe A compartment is included to store associated cables and terminations See Figure 4 Figure 4 Equipment locations in th...

Страница 26: ...must be at least 500 MHz When using a TCP305A Current Probe the oscilloscope bandwidth must be at least 250 MHz After you have connected the amplifier to the oscilloscope allow the equipment to warm...

Страница 27: ...be while the probe is clamped around a live conductor or the probe may suffer electrical damage If you connect a probe to the wrong amplifier for example a TCP312A to a TCPA400 the NONCOMPATIBLE PROBE...

Страница 28: ...fier will light The current probes can be used to measure current on uninsulated wires However the circuit must be de energized when connecting or removing the current probe The slide operation of the...

Страница 29: ...the lock button 2 Squeeze the handle until the core is open 3 Place the probe core around the conductor Figure 7 Unlock and open the TCP303 and TCP404XL 4 To lock the probe release the squeeze handle...

Страница 30: ...nal magnetic field Periodically during normal use To degauss and autobalance a current probe perform these steps 1 Verify that the current probe is connected to the amplifier 2 Remove the current prob...

Страница 31: ...he amplifier PROBE DEGAUSS AUTOBALANCE button The NOT TERMINATED INTO 50 LED is lighted if impedance is not 50 If this is the case make necessary changes For example use a 50 termination 5 After the d...

Страница 32: ...place the probe around the conductor under test and then lock the slide 2 For correct measurement polarity make sure the probe arrow is pointing in the direction of conventional positive to negative...

Страница 33: ...amplifier PROBE DEGAUSS AUTOBALANCE button WARNING The current probes can be used to measure current on uninsulated wires However the circuit must be de energized when connecting or removing the curre...

Страница 34: ...Getting Started 16 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...

Страница 35: ...l magnetism from the attached current probe A multi color LED indicates the status of the degauss circuit 2 The MANUAL BALANCE buttons allow you to fine adjust DC offset from the amplifier The adjacen...

Страница 36: ...button blinks red whenever the amplifier detects that the current probe needs degaussing The amplifier cannot detect all circumstances that require probe degaussing so you may need to degauss the pro...

Страница 37: ...ope input is set to 50 impedance Proper cabling is shown See Figure 1 on page 1 If your oscilloscope does not have 50 impedance settings for inputs you can place a 50 feed through termination on the o...

Страница 38: ...his button to power on the amplifier When the amplifier is in STANDBY mode the amplifier is in a limited power mode Most of the secondary circuitry is disabled but the line voltage remains connected t...

Страница 39: ...e input impedance of your oscilloscope must be 50 Make sure your amplifier OUTPUT is connected to an oscilloscope input using a 50 BNC cable and that the oscilloscope input is set to 50 impedance The...

Страница 40: ...t Remember altering the DC Gain Adjustment may cause the probe to not meet the warranted DC Accuracy specification GPIB Operation The TCPA300 and TCPA400 Current Probe Amplifiers do not have direct GP...

Страница 41: ...ill cause a residual offset in the current probe and the amplifier may report an inaccurate result If the impedance of your circuit is higher than that shown the degauss procedure succeeds because the...

Страница 42: ...onductor Note that lacquer coatings like those typically found on transformer windings do not provide sufficient reliable insulation for use with current probes The lacquer coating can be easily nicke...

Страница 43: ...h the conventional current flow I1 in the direction of the probe arrow is carrying the greater current Conventional current flows from positive to negative 4 To adjust for a current null adjust the cu...

Страница 44: ...e Figure 13 1 A low frequency square wave using AC coupling The signal exhibits low frequency rolloff 2 By changing the amplifier coupling to DC the pulse is displayed as truly square as shown Figure...

Страница 45: ...ely measure regardless of how short within bandwidth limitations the pulse duration is Ampere Second Product defines the maximum width of pulsed current that you can measure when the pulse amplitude i...

Страница 46: ...lowable pulse width PWmax For example the TCP312A Current Probe has a maximum Ampere second product of 500 A s in the 10 A V range setting If a pulse measured with a TCP312A has a peak current of 40 A...

Страница 47: ...by the pulse width The quotient is the maximum allowable pulse amplitude the peak amplitude of the measured pulse must be less than this value For example the TCP312A Current Probe has a maximum Amper...

Страница 48: ...the noncontinuous current defined as measurement time and the measurement period A continuous and noncontinuous current is shown and how to calculate the duty cycle See Figure 15 Figure 15 Duty cycle...

Страница 49: ...Figure 36 on page 58 Compare the maximum measurement time allowed for a duty cycle of 20 At 750 A you have 3 minutes of safe measurement time versus 17 minutes for a smaller noncontinuous current ampl...

Страница 50: ...DC current range of your probe you can add offset bucking current with a second conductor WARNING To avoid injury or loss of life from shock or fire do not put more than one uninsulated conductor at...

Страница 51: ...g current is equal to the current flowing in the conductor multiplied by the number of turns wound around the probe For example if the second conductor has a current of 100 mA DC and is wrapped around...

Страница 52: ...ned as any conductor without insulation or without insulation rated for the voltage present on the conductor under test When viewing the signal on the oscilloscope screen divide the displayed amplitud...

Страница 53: ...de is shorted or open See Figure 20 on page 36 A TCP303 Current Probe was used for this high current low voltage application The TCP303 is degaussed and clamped around the positive battery lead from t...

Страница 54: ...oximately 23 A and a maximum peak of approximately 31 A The waveform shows a continuous cycle with no dropouts so the alternator circuit appears to be functioning properly A single phase diode failure...

Страница 55: ...nected to an inductor that has low resistance See Figure 21 1 Connect the inductor across the output terminals of the pulse generator 2 Maintain a constant voltage across the inductor 3 Clamp the curr...

Страница 56: ...shown here Figure 22 Linear current vs time ramp 5 Calculate the inductance using the following formula where L is the inductance in henries E is the voltage of the pulse generator dt is the change i...

Страница 57: ...time required for the current to rise or fall 63 2 of the total current value and R is the source resistance of the pulse generator Figure 23 High impedance source current ramp Continuity Test of Mult...

Страница 58: ...io of coil current to input current The accuracy of this method is limited by the current measurement accuracy The following method allows more precise turns measurement For a more precise turns count...

Страница 59: ...measurement system into a sophisticated analysis tool that quickly measures and analyzes many different applications such as Power dissipation at switching devices and magnetic components in switchin...

Страница 60: ...ases 15 minutes is sufficient All LEDs are flashing All LEDs flashing indicates a thermal shutdown Power cycle the system and let it cool before taking measurements In most cases 15 minutes is suffici...

Страница 61: ...tton The Current Probe has been overloaded dropped or exposed to magnetic field You can use the MANUAL BALANCE buttons to null out any residual DC offset after performing a Probe Degauss Autobalance r...

Страница 62: ...accuracy Check that the probe slider is fully closed and locked Check and clean the probe transformer and lid Degauss the probe The amplifier or the oscilloscope may be AC coupled Set the COUPLING to...

Страница 63: ...being displayed instead of normal conditions 3 The four fault indicator LEDs above the ON STANDBY switch are used to form a four bit binary error code See Figure 27 on page 46 The error codes are def...

Страница 64: ...error occurred while nulling out the internal DC offset of the amplifier Power cycle the amplifier and run the Degauss Autobalance adjustment routine again If the error reoccurs then the amplifier nee...

Страница 65: ...r needs service 15 Unspecified error internal software error Power cycle the amplifier and run the Degauss Autobalance adjustment routine again If the error reoccurs then the amplifier needs service C...

Страница 66: ...Troubleshooting and Error Codes 48 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...

Страница 67: ...output is properly terminated into 50 Specifications are separated into two categories warranted specifications and nominal or typical characteristics Warranted Specifications Warranted specification...

Страница 68: ...Limit measurement bandwidth to 20 MHz Signal Delay typical 1 approximately 17 ns approximately 19 ns approximately 53 ns approximately 103 ns Maximum Bare Wire Voltage 150 CAT II bare wire 300 V CAT...

Страница 69: ...47 Hz to 440 Hz single phase Maximum Power 50 Watts maximum Fuse Rating 3 15 AH 250 V Not operator replaceable 1 Includes probe amplifier and TEKPROBE to TEKPROBE interface cable or 012 0117 xx BNC ca...

Страница 70: ...Figure 28 Probe jaw dimensions nominal Table 12 Shipping weights and dimensions TCPA300 and TCPA400 Amplifiers TCP305A and TCP312A Probes TCP303 Probe TCP404XL Probe Dimensions Length 298 5 mm 11 75...

Страница 71: ...onoperating 40 C to 75 C 40 F to 167 F Humidity Operating 5 to 95 R H to 30 C 86 F 5 to 85 R H 30 C to 50 C 86 F to 122 F Nonoperating 5 to 95 R H to 30 C 86 F 5 to 85 R H 30 C to 75 C 86 F to 167 F A...

Страница 72: ...ions Performance Graphs Figure 29 Frequency derating TCP312A Figure 30 Frequency derating TCP305A Figure 31 Frequency derating TCP303 54 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes Us...

Страница 73: ...Specifications Figure 32 Frequency derating TCP404XL TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual 55...

Страница 74: ...Specifications Figure 33 Insertion impedance versus frequency 56 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...

Страница 75: ...Specifications Figure 34 Specified operating area of the probes TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual 57...

Страница 76: ...ing three graphs Details about the relationship between measurement factors are available See page 30 Measuring Noncontinuous Current with the TCP404XL Probe Figure 35 Measuring 750A noncontinuous at...

Страница 77: ...23 degrees ambient temperature 600 A can be measured continuously with the TCP404XL probe Emissions which exceed the levels required by this standard may occur when this equipment is connected to a te...

Страница 78: ...Specifications 60 TCPA300 400 Amplifiers and TCP300A 400 Series Current Probes User Manual...

Страница 79: ...ive to negative The arrows on current probes point in the direction of conventional current flow Conventional current flow is in the opposite direction of electron current flow degauss A process of el...

Страница 80: ...ircuit when the current probe is clamped around a test conductor magnetic susceptibility A figure expressing the amount of current induced into the probe by an external magnetic field of known intensi...

Страница 81: ...See also conventional current Error codes correcting 47 Error Codes displaying 45 shutdown error 47 Extending the current range of the probes 32 F flux 61 Frequency derating 27 Front panel controls 17...

Страница 82: ...tizing 12 18 description 2 extending the current range 32 increasing the current limit 32 maximum current limits 27 operating the probe slide 10 saturation 27 R RANGE button and indicator 20 S Safety...

Отзывы: