Adjustments
DC Gain Accuracy
This procedure describes the DC gain accuracy adjustments for the probe at the 5 A and 30 A range settings.
Adjust the 5 A Range
1.
Complete steps 1 through 18 of the
DC
Gain Accuracy Test
in the
Performance
Veri
fi
cation
procedure. (See page 54,
Gain Accuracy
2.
Calculate the waveform target value (g)
using the calculated oscilloscope gain
accuracy value (b) from step 7 of the
performance test, and the formula shown.
3.
Adjust the 5 A DC gain control in the probe
compensation box to the calculated target
value from step 2, ±0.02A.
64
TCP0030 Current Probe Instruction Manual