• CISPR 11. Radiated and conducted emissions, Group 1, Class A
• IEC 61000-4-2. Electrostatic discharge immunity
• IEC 61000-4-3. RF electromagnetic field immunity
• IEC 61000-4-4. Electrical fast transient / burst immunity
• IEC 61000-4-5. Power line surge immunity
• IEC 61000-4-6. Conducted RF immunity
• IEC 61000-4-11. Voltage dips and interruptions immunity
Australia/New Zealand Declaration of Conformity - EMC
Complies with the EMC provision of the Radiocommunications Act per the following standard, in accordance with ACMA:
• EN 61326-1. Radiated and conducted emissions, Group 1, Class A
• EN 61326-1 and EN 61326-2-1. Radiated and conducted emissions, Group 1, Class A
Safety compliance
This section lists the other safety compliance information.
Equipment type
Test and measuring equipment.
Safety class
Class 1 – grounded product.
Pollution degree description
A measure of the contaminants that could occur in the environment around and within a product. Typically the internal environment inside a
product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
• Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category are generally encapsulated,
hermetically sealed, or located in clean rooms.
• Pollution Degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by
condensation must be expected. This location is a typical office/home environment. Temporary condensation occurs only when the
product is out of service.
• Pollution Degree 3. Conductive pollution, or dry, nonconductive pollution that becomes conductive due to condensation. These are
sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct
wind.
• Pollution Degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical outdoor locations.
Pollution degree rating
Pollution Degree 2 (as defined in IEC 61010-1). Rated for indoor, dry location use only.
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
3
For compliance with the EMC standards listed here, high quality shielded interface cables that incorporate low impedance connection between the cable shield and
the connector shell should be used.
4
Equipment may not meet the immunity requirements of applicable listed standards when test leads and/or test probes are connected due to coupling of
electromagnetic interference onto those leads/probes. To minimize the influence of electromagnetic interference, minimize the loop area between the unshielded
portions of signal and associated return leads, and keep leads as far away as possible from electromagnetic disturbance sources. Twisting unshielded test leads
together is an effective way to reduce loop area. For probes, keep the ground return lead as short as possible and close to the probe body. Some probes have
accessory probe tip adapters to accomplish this most effectively. In all cases, observe all safety instructions for the probes or leads used.
Compliance information
6
Содержание TBS2000B Series
Страница 12: ...Tektronix CAT 1 1 III IV 2...
Страница 13: ...CAT BNC 3...
Страница 14: ...4...
Страница 18: ...1 2 3 Power 1 Power 2 8...
Страница 19: ...1 1 2 3 4 TekVPI 2 PROBE COMP 1 kHz 5 V 3 USB 2 0 9...
Страница 20: ...1 2 LAN 10 100 Base T 3 USB 2 0 USB Tektronix TBS2000B USB Wi Fi U 4 USB 2 0 USB 2 0 PC 10...
Страница 22: ...2...
Страница 23: ...CAT 1 1 III IV 3...
Страница 24: ...CAT 2 2 BNC 4...
Страница 25: ...DANGER 5...
Страница 27: ...1 1 2 3 4 2 IEC 61010 1 IP IP20 IEC 60529 1 II III IV II 7...
Страница 29: ...3 1 2 AC 3 1 2 9...
Страница 30: ...1 1 2 3 4 TekVPI 2 PROBE COMP 5 V 1 kHz 3 USB 2 0 10...