background image

3. If steps

1

 on page 13 and

2

 on page 13 do not remove all the dust or dirt, all modules in the oscilloscope except the Display Module and

Front Panel Module may be spray washed using a solution of 75% isopropyl alcohol.

• During cleaning of the Acquisition board (Main board), make sure that no residue is left in the attenuators. Debris or residue can

cause several problems with signal fidelity that cannot be corrected by calibration. When handling the board, avoid handling the
attenuators since skin oils may cause problems.

• Extensive cleaning of the Display Module is difficult because water must not be trapped between the display shield and the display.

The polarizers integrated into the display degrade with long-term exposure to water. The display shield and display must be
replaced as an assembly if significant moisture or debris become trapped inside. Cotton tipped applicators and lint-free rags
moistened in isopropyl alcohol and/or deionized water offer the best cleaning with the least opportunity for damage to the
assembly.

• The Front-Panel module can be damaged by deposition of film or debris inside the knob encoders. To avoid this, be careful not to

drip liquid or blow debris into the encoders.

4. Spray wash dirty parts with the 75% isopropyl alcohol solution and wait 60 seconds for the majority of the alcohol to evaporate.
5. Use hot (49 °C to 60 °C / 120 °F to 140 °F) deionized water to thoroughly rinse the parts.
6. Dry all parts with low-pressure, deionized air.
7. Dry all components and assemblies in an oven or drying compartment using low-temperature (52 °C to 66 °C / 125 °F to 150 °F)

circulating air.

Lubrication

Periodic lubrication is not required for the oscilloscope.

Maintenance

TBS1000C Series Digital Storage Oscilloscope Service Manual

14

Содержание TBS1000C Series

Страница 1: ...TBS1000C Series Digital Storage Oscilloscope Service Manual P077158401 077 1584 01...

Страница 2: ......

Страница 3: ...TBS1000C Series Digital Storage Oscilloscope Service Manual Register now Click the following link to protect your product www tek com register 077 1584 01...

Страница 4: ...ued and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of...

Страница 5: ...ion and cleaning 12 General care 12 Interior cleaning 12 Exterior cleaning 12 Exterior inspection 12 Interior inspection 13 Cleaning procedure interior 13 Lubrication 14 Adjustments 15 Replaceable par...

Страница 6: ...t 13 Table 2 Internal inspection checklist 13 Table 3 Replaceable parts list 17 Table 4 Replaceable parts LCD Front panel Front case Power supply Main board and Rear case 18 List of Tables TBS1000C Se...

Страница 7: ...and certified for the country of use Ground the product This product is grounded through the grounding conductor of the power cord To avoid electric shock the grounding conductor must be connected to...

Страница 8: ...als before you clean the product Provide proper ventilation Refer to the installation instructions in the manual for details on installing the product so it has proper ventilation Slots and openings a...

Страница 9: ...ice procedures Read this Service safety summary and the General safety summary before performing any service procedures To avoid electric shock Do not touch exposed connections Do not service alone Do...

Страница 10: ...ut the nature of the potential hazards and any actions which have to be taken to avoid them This symbol may also be used to refer the user to ratings in the manual The following symbols may appear on...

Страница 11: ...al available on the Tektronix Web site Go to www tektronix com manuals and search for the user manual for your product available in 11 languages Programmer Commands The oscilloscope programmer manual...

Страница 12: ...and disconnect it from line voltage before performing any procedure that follows Exterior cleaning CAUTION Avoid the use of chemical cleaning agents which might damage the plastics used in this oscill...

Страница 13: ...oscilloscope Table 2 Internal inspection checklist Item Inspect for Repair action Circuit boards Loose broken or corroded solder connections Burned circuit boards Burned broken or cracked circuit run...

Страница 14: ...replaced as an assembly if significant moisture or debris become trapped inside Cotton tipped applicators and lint free rags moistened in isopropyl alcohol and or deionized water offer the best clean...

Страница 15: ...test equipment signal sources and measuring equipment to verify the performance of the instrument To determine if an adjustment is required use the Performance Verification procedure in the TBS1000C S...

Страница 16: ...ision level of the part For some parts in this manual you will find the letters XX in place of the revision level number When you order parts Tektronix will provide you with the most current part for...

Страница 17: ...arated from the description by a colon Because of space limitations an item name may sometimes appear as incomplete Use the U S Federal Catalog handbook H6 1 for further item name identification Abbre...

Страница 18: ...panel board 8 131 6417 XX 1 Contact Electrical 9 220 0265 XX 3 BNC nuts 10 210 1039 XX 3 BNC locking washer 11 441 2909 XX 1 Chassis ASSY 12 211 1273 XX 6 Screw Threaded 13 870 1704 XX 1 Main PCB ass...

Отзывы: