Performance veri
fi
cation
8.
Set oscilloscope probe offset to 0.0 V.
9.
Set the DMM to DCV.
10.
Set the DC voltage source to +1.00 VDC
and enable the output.
11.
Record the DMM measurement in the test
record.
12.
Set the DC voltage source to -1.00 VDC.
13.
Record the DMM measurement in the test
record.
NOTE.
An unacceptable error value may
result if a precision 50
Ω
termination is not
used for the recommended termination, or if
the oscilloscope probe offset is not set to zero.
46
TAP1500 1.5 GHz, 10X Active Probe Instruction Manual