Performance Verification
RTX130B QAM & VSB RF Signal Generator Service Manual
3-9
This test verifies the internal clock output level and frequency accuracy.
Equipment
i d
Oscilloscope
q p
required
Frequency counter
50
Ω
BNC cable
1.
Use the 50
Ω
BNC cable to connect the Trig In/Out connector on the
RTX130B to the oscilloscope CH1 input. See Figure 3-6.
Oscilloscope (TDS5104B)
50
Ω
BNC cable
RTX130B rear panel
Figure 3-6: Equipment connection for checking the internal clock output level
2.
Select
Play > Others
on the RTX130B to open the
Others
dialog box.
3.
In the dialog box, select the
Ext Trigger BNC
button to open the Ext
Trigger BNC dialog box.
4.
In the dialog box, set
BNC IN/OUT
to
Output
and
BNC OUT Selection
to
27 MHz
.
Internal Clock Output
Level and Frequency
Accuracy
Содержание RTX130B
Страница 4: ......
Страница 14: ...Service Safety Summary x RTX130B QAM VSB RF Signal Generator Service Manual...
Страница 16: ...Environmental Considerations xii RTX130B QAM VSB RF Signal Generator Service Manual...
Страница 21: ...Specifications...
Страница 22: ......
Страница 38: ...Specifications 1 16 RTX130B QAM VSB RF Signal Generator Service Manual...
Страница 39: ...Theory of Operation...
Страница 40: ......
Страница 48: ...Theory of Operation 2 8 RTX130B QAM VSB RF Signal Generator Service Manual...
Страница 49: ...Performance Verification...
Страница 50: ......
Страница 102: ...Performance Verification 3 52 RTX130B QAM VSB RF Signal Generator Service Manual...
Страница 103: ...Maintenance...
Страница 104: ......
Страница 153: ...Options...
Страница 154: ......
Страница 156: ...Options 5 2 RTX130B QAM VSB RF Signal Generator Service Manual...
Страница 157: ...Diagrams...
Страница 158: ......
Страница 160: ...Diagrams 6 2 RTX130B QAM VSB RF Signal Generator Service Manual...
Страница 162: ...6 4 RTX130B QAM VSB RF Signal Generator Service Manual...
Страница 164: ...6 6 RTX130B QAM VSB RF Signal Generator Service Manual...
Страница 165: ...Replaceable Parts List...
Страница 166: ......