Functional Check and Calibration
Calibrate the Probe
You must calibrate the probe once on each
oscilloscope channel that you intend to use.
The probe calibration data is then stored for
future use on those channels.
1.
Connect the probe to any channel (1–4) of
the oscilloscope.
2.
Set the oscilloscope to display the channel.
3.
Connect the probe A and B inputs to the A
and B inputs on the TriMode DC Calibration
board, using the connectors that mate to
the adapter/tip you are using.
4.
Connect a BNC cable from the DC Probe
Cal output connector on the oscilloscope
to the BNC connector on the TriMode DC
Calibration board.
5.
Connect the cable from the TekConnect
control box adapter to the connector on the
calibration board.
6.
Connect the control box to the Aux
In channel on the oscilloscope. Allow
the
fi
xture and probe to warm up for
20 minutes.
18
P7600 Series TriMode Probes Quick Start User Manual