Functional Check
16
P7225 2.5 GHz 10X Active Probe User Manual
3.
Using a ground lead and SMT KlipChip, connect the probe
ground to the instrument ground on the host instrument.
4.
Using a standard tip, hold the probe to the center conductor of the
PROBE COMPENSATION output on the host instrument.
5.
Press
AUTOSET
, or adjust the host instrument to display a stable
calibration waveform. (You may need to adjust the probe offset to
display the waveform.)
NOTE
. Now is a good time to perform a probe calibration routine
from your instrument. Refer to the online help of your host instrument
for calibration routine procedures.
6.
Disconnect the probe tip from the host instrument, and ground the
probe tip. (Connect the KlipChip to the probe tip.)
7.
Set the probe offset to 0.0 V to make the host instrument display
at the ground reference.
8.
Adjust the V/div to keep the waveform on the screen.