Maintenance
7
P6562A Instructions
1. Remove the top plastic cover from the compensation box. (To
remove the cover, gently insert a tool, such as a small flat-blade
screwdriver, under one side of the cover and carefully pry up and
out.)
2. Connect the P6562A to the oscilloscope.
3. On the calibration generator, change the pulse period to 1
s
(1 MHz).
4. Connect the probe tip to the calibration signal. If you are using a
PG 506A, first connect a feed-through termination to the fast-rise
output, then connect the probe tip to the feed-through termina-
tion. The signal-ground clip can be used for the ground
connection.
5. Adjust the oscilloscope to view the first 40 ns of the waveform.
See figure 4.
6. Alternately adjust HF1 and HF2 for overall flatness. (These
adjustments interact.)
7. Adjust HF3 for the squarest waveform.