Operating Basics
3.
If the retention assembly has minimal play and you cannot see a gap between
the bottom of the assembly and the circuit board surface, then move the probe
with bad data from one logic analyzer probe location to another.
4.
If the problem follows the probe, then the probe is the problem. Visually
inspect the cLGA interface clip on the probe for any damage or missing
c-spring metal contacts.
If there is damage to the interface clip, or if any c-spring metal contacts are
missing, replace the cLGA interface clip. (See page 27,
5.
If the problem doesn't follow the probe, it is either the logic analyzer or the
probe connection at its previous location. Move the probe back to the original
location to be certain it was not a connection problem at the logic analyzer end.
6.
Place another probe in the retention assembly of the original probe. If the new
probe acquires data, then the old probe is probably at fault.
10
P6450 High-Density Logic Analyzer Probe Instruction Manual
Содержание P6450
Страница 4: ......
Страница 12: ...Compliance Information viii P6450 High Density Logic Analyzer Probe Instruction Manual...
Страница 16: ...Preface xii P6450 High Density Logic Analyzer Probe Instruction Manual...
Страница 42: ...Specifications 26 P6450 High Density Logic Analyzer Probe Instruction Manual...
Страница 46: ...Maintenance 30 P6450 High Density Logic Analyzer Probe Instruction Manual...
Страница 50: ...Replaceable Parts Figure 21 Optional accessories 34 P6450 High Density Logic Analyzer Probe Instruction Manual...