Adjustments
P6245 Instruction Manual
4- 11
Offset Range
Use this procedure to adjust the offset range of the probe. The offset
zero of the probe should be checked before making any adjustment
to the offset range.
The offset range of the probe is approximately --10 to +10 volts. The
offset range is not warranted; however, this adjustment procedure is
provided for your convenience.
NOTE
. If you are using an 1103 Probe Power Supply, make sure that
the oscilloscope is set to 50
Ω
termination with DC coupling.
See Figure 4--4 for the location of the offset range adjustment.
1.
Set the power supply (using the DMM) to +10.00 VDC. (If
10.00 V cannot be set exactly, set the power supply just below
10.00 V.)
2.
Connect the probe to the power supply as shown in Figure 4--3.
3.
Set the channel’s vertical offset to match the DMM reading.
4.
Set the channel’s vertical range to 0.1V/div.
5.
Observe the DC value of the displayed trace.
6.
Adjust
Offset Range
for an oscilloscope reading that matches the
DMM reading.
NOTE
. After the offset zero and offset range have been adjusted,
perform a probe calibration and functional check to verify perfor-
mance.
Содержание P6245
Страница 2: ......
Страница 3: ...P6245 1 5 GHz 10X Active Probe ZZZ Instruction Manual xx www tektronix com 070 8995 05...
Страница 6: ......
Страница 12: ...Manual Organization vi P6245 Instruction Manual...
Страница 13: ...Getting Started...
Страница 14: ......
Страница 25: ...Operating Basics...
Страница 26: ......
Страница 30: ...Operating Basics 2 4 P6245 Instruction Manual...
Страница 35: ...Specifications...
Страница 36: ......
Страница 43: ...Fran ais...
Страница 44: ......
Страница 45: ...P6245 Sonde active 10X 1 5 GHz ZZZ Manuel d utilisation xx www tektronix com 070 8995 05...
Страница 48: ......
Страница 54: ...Structure du manuel vi Manuel d utilisation P6245...
Страница 55: ...Mise en route...
Страница 56: ......
Страница 64: ...Caract ristiques et accessoires 1 8 Manuel d utilisation P6245...
Страница 69: ...Fonctionnement...
Страница 70: ......
Страница 74: ...Fonctionnement 2 4 Manuel d utilisation P6245...
Страница 79: ...Caract ristiques techniques...
Страница 80: ......
Страница 87: ...Deutsch...
Страница 88: ......
Страница 89: ...P6245 1 5 GHz 10X Activer Tastkopf ZZZ Benutzerhandbuch xx www tektronix com 070 8995 05...
Страница 92: ......
Страница 98: ...Aufbau des Handbuchs vi P6245 Benutzerhandbuch...
Страница 99: ...Inbetriebnahme...
Страница 100: ......
Страница 113: ...Grundlegendes zum Betrieb...
Страница 114: ......
Страница 118: ...Grundlegendes zum Betrieb 2 4 P6245 Benutzerhandbuch...
Страница 123: ...Spezifikationen...
Страница 124: ......
Страница 131: ...Espa ol...
Страница 132: ......
Страница 133: ...P6245 Sonda Activa de 1 5 GHz y 10X ZZZ Manual de Instrucciones xx www tektronix com 070 8995 05...
Страница 136: ......
Страница 142: ...Estructura del Manual vi P6245 Manual de Instrucciones...
Страница 143: ...Puesta en Marcha...
Страница 144: ......
Страница 157: ...Instrucciones B sicas de Manejo...
Страница 158: ......
Страница 162: ...Instrucciones B sicas de Manejo 2 4 P6245 Manual de Instrucciones...
Страница 167: ...Especificaciones...
Страница 168: ......
Страница 175: ...Service Information...
Страница 176: ......
Страница 180: ......
Страница 184: ...Theory of Operation 4 4 P6245 Instruction Manual...
Страница 192: ...Adjustments 4 12 P6245 Instruction Manual...
Страница 198: ...Maintenance 4 18 P6245 Instruction Manual...
Страница 202: ...Troubleshooting 4 22 P6245 Instruction Manual...
Страница 203: ...Replaceable Parts...
Страница 204: ......
Страница 208: ...Replaceable Parts P6245 Instruction Manual 5 4 1 3 2 3 Figure 5 1 P6245 Replaceable Parts...
Страница 210: ...Replaceable Parts P6245 Instruction Manual 5 6 1 2 3 4 5 6 Figure 5 2 P6245 Standard Accessories...
Страница 212: ...Replaceable Parts P6245 Instruction Manual 5 8 1 2 3 4 5 Figure 5 3 P6245 Optional Accessories...
Страница 214: ...Replaceable Parts P6245 Instruction Manual 5 10...