Built-In Security Features
To Overwrite Protected
User Data
This procedure overwrites everything that is stored in the protected user data area.
Send the following commands to the oscilloscope:
:PASSWORD “XYZZY”
(or current password if changed from the default of
“XYZZY”)
:NEWPASS “ANYTHING”
:PASSWORD “ANYTHING”
*PUD #3300xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx[repeat
for a total of 300 x’s]
NOTE.
If you do not have access to a program that supports sending programmatic
commands to the instrument, copy the preceding commands to a text
fi
le that ends
in “.set” and recall the
fi
le from the Recall Setup menu.
For more information on using programming commands, refer to the Programmer
Manual for this instrument, available at www.tektronix.com/downloads.
10
Declassi
fi
cation and Security Instructions