Model DMM7510 7½ Digit Multimeter Application Manual
Section 7: Capturing and analyzing waveforms
DMM7510-904-01 Rev. D March 2021
7-9
Run the test using the front panel
To run the duty-cycle test from the front panel:
1. Press the
POWER
button on the front panel to turn on the instrument.
2. On the FUNCTIONS swipe screen, select
Digi V
to select the digitize voltage function.
3. Swipe to display the SETTINGS swipe screen.
4. Set the Sample Rate to
1,000,000
.
5. Set the Count to
100
.
6. Swipe to the
GRAPH
swipe screen.
7. Select the graph icon in the upper right of the swipe bar to view the graph in the full screen.
8. Swipe the bottom of the graph screen until the Cursor swipe bar is displayed.
9. Select the button on the Cursor swipe bar until
Vert
is displayed. Vertical cursors are displayed
on the graph.
10. Swipe the vertical cursor lines to move them to the changes in time of a single cycle. The change
is displayed as
X in the Cursor swipe bar.
Test results
The front-panel graph shows results that are similar to the following.
Figure 15: Duty cycle 1