Run the test using the front panel
To run the test using the front panel:
1. Press the
POWER
button on the front panel to turn on the instrument.
2. On the FUNCTIONS swipe screen, select
Digi V
to select the Digitize Voltage measure function.
3. Swipe to display the
SETTINGS
swipe screen.
4. Set the Sample Rate to
1,000,000
.
5. Set the Count to
100
.
6. Swipe to display the
GRAPH
swipe screen.
7. Select the graph icon on the upper right of the swipe screen to open the Graph screen.
8. To analyze the voltage ripple peak-to-peak value, swipe the bottom of the graph screen until the
Cursor Enable button is displayed.
9. Select the
Cursor Enable
button until Horiz is displayed. The horizontal cursors are displayed.
10. Touch the yellow horizontal cursor lines on the graph and move them to measure the
peak-to-peak ripple amplitude.
11. The
Y value is shown on the graph swipe bar.
Figure 13: Results of the ripple noise test