Section 3: Adjustment
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
3-22
DMM6500-905-01 Rev. E June 2021
Command parameters for a dc current adjustment
When calibrating your DMM6500 for a dc current adjustment, use the following command parameters.
Send each command parameter twice. First, send the parameter using the
setup
command. Second,
send the parameter using the
execute
command along with the calibrator stimulus value.
The example code below uses the first line of code from the table below. Send all command
parameters using this command.
cal.adjust.step.setup("cal_DCI_10uA_fs_pos")
cal.adjust.step.execute("cal_DCI_10uA_fs_pos", <
calibrator stimulus value
>)
You must send the
cal_TS6
first.