DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Section 2: Performance verification
DMM6500-905-01 Rev. E June 2021
2-29
Figure 10
: Connections for 100 MΩ verification
2. Set the calibrator for 2-wire resistance with external sense off.
3. On the DMM6500, press the
FUNCTION
key and select
2W Res
.
4. On the home screen, select the button next to Range and select
100 MΩ
.
5. Source the nominal full-
scale resistance values for the 100 MΩ range as shown in the f
ollowing
table.
The values and limits in the following tables are for example only. You must calculate test limits
based on the actual resistance values output by your calibrator or resistance source (see
(on page 2-4)).
Range (Ω)
Nominal calibrator
values (Ω)
Actual calibrator
(Ω)
Lower limit (Ω)
Upper limit (Ω)
100 M
0
0
-10
10
100
100.001
99.798
100.204