Section 7: TSP command reference
Models 707B and 708B Switching Matrix Reference Manual
7-224
707B-901-01 Rev. B / January 2015
tsplink.trigger[N].reset()
This function resets some of the TSP-Link trigger attributes to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.trigger[
N
].reset()
N
The trigger line (1 to 3)
Details
The
tsplink.trigger[
N
].reset()
function resets the following attributes to their factory
defaults:
•
tsplink.trigger[
N
].mode
•
tsplink.trigger[
N
].stimulus
•
tsplink.trigger[
N
].pulsewidth
This also clears
tsplink.trigger[
N
].overrun
.
Example
tsplink.trigger[3].reset()
Resets TSP-Link trigger line 3
attributes back to factory default
values.
Also see
(on page 7-220)
(on page 7-222)
(on page 7-223)
tsplink.trigger[N].stimulus
This attribute specifies the event that causes the synchronization line to assert a trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Reset
Recall setup
TSP-Link trigger N reset
Create
configuration script
0
Usage
eventID
= tsplink.trigger[
N
].stimulus
tsplink.trigger[
N
].stimulus =
eventID
eventID
The event identifier for the triggering event
N