Model 6517B Electrometer User's Manual
Section 4: Basic measurements
6517B-900-01 Rev. B August 2022
4-19
To measure resistivity:
1. Press the
Z-CHK
key to enable zero check (
ZeroCheck
displayed).
2. Select the
R
key to select the resistance function.
3. Press the
CONFIG
key.
4. Press the
R
key. The resistance configuration menu is displayed.
5. Select
MEAS-TYPE
. Press the
ENTER
key.
6. Select
RESISTIVITY
. Press the
ENTER
key.
7. Select
SURFACE
or
VOLUME
. Press the
ENTER
key.
8. If you selected VOLUME:
a. Select
THICKNESS
. Press the
ENTER
key.
b. Set the thickness. Press the
ENTER
key.
9. Select
FIXTURE-MODEL
. Press the
ENTER
key.
10. Select
MODEL-8009
. Press the
ENTER
key.
11. Press the
EXIT
key twice to return to the CONFIGURE OHMS menu.
12. Select
VSOURCE
. Press the
ENTER
key.
13. Select AUTO or MANUAL:
AUTO:
The instrument automatically selects the optimum voltage source value (40 V or
400 V) for the measurement range.
MANUAL:
You select the voltage source range and value.
14. Press the
EXIT
key to return to the measurement display.
15. If VSOURCE is set to MANUAL:
a. Use the cursor keys and the
VOLTAGE SOURCE
adjustment keys to set the voltage level.
b. To change the voltage source range, press the
CONFIG
key and then the
OPER
key, then select
RANGE
.
16. Connect the sample to be measured to the 6517B, as shown in the following figure.
Содержание Keithley 6517B
Страница 2: ...Electrometer User s Manual Model 6517B ...