To enable METER-CONNECT:
1. Press the
CONFIG
key.
2. Press the
OPER
key. The CONFIG V-SOURCE menu is displayed.
3. Select
METER-CONNECT
. Press the
ENTER
key.
4. Select
ON
to enable the internal voltage source LO to ammeter LO connection. Press the
ENTER
key.
The maximum common-mode input voltage (the voltage between input low and chassis
ground) is 500 V
PEAK
. To prevent electrical shock that could cause injury or death, do
not exceed this value.
Voltage source output connections
The voltage source output is accessed at the V SOURCE OUT LO and HI binding posts on the rear
panel, as shown in the following figure.
Figure 15: Volage source output basic connections
Using these terminals places the independent voltage source in series with the external circuit (R
L
),
as shown in the following figure.