FAQs
How should I clean and use the front-panel display?
If you need to clean the front-panel LCD touchscreen display,
use a soft dry cloth. If necessary, use a microfiber cloth
dampened with an ammonia-free glass cleaner. Do not spray
cleaning fluids onto the display. You can also use a mixture of
70% isopropyl alcohol and 30% water.
Do not use sharp objects, such as a screwdriver, pen, or
pencil, to touch the touchscreen. It is strongly recommended
to only use fingers to operate the instrument. Use of
clean-room gloves is supported for the touchscreen.
My data looks odd or is wrong. What should I do?
Verify the connections from the instrument to the test fixture.
Also, check the connections from the DUT to the test fixture
socket.
I cannot unplug the mini-triaxial cable (4200A-MTRX)
from the SMU. What should I do?
The mini-triaxial connector is a locking connector. To remove
it, pull the knurled part of the connector back.
A configuration error has been detected and I cannot
launch Clarius. What should I do?
This occurs when the physical configuration does not
match the configuration defined in KCon or when there are
communication problems between instruments and the
4200A-SCS.
The error can also occur if the preamplifier or RPM is
removed or reconnected. Note that preamplifiers are SMU
specific. For example, a preamplifier that is configured for
SMU1 cannot be connected to SMU2.
To verify system configuration
:
1. Run KCon.
2. Select
Validate
.
To update the system configuration
:
1. Run KCon.
2. Select
Update
.