![Tektronix KEITHLEY 4200A-SCS-PKC Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/tektronix/keithley-4200a-scs-pkc/keithley-4200a-scs-pkc_quick-start-manual_1078026019.webp)
I-V Test
To configure the test:
1. Select
Configure
to open the pane shown here. This is
where you define the test.
2. The
vds-id
settings should be:
• SMU3: Provide a voltage step function for four
different gate voltages (2 V, 3 V, 4 V ,and 5 V).
• SMU2: Perform a 51-point sweep of drain voltage (0 V
to 5 V) at each gate voltage. A current measurement
is made at each voltage sweep point.
3. Change the settings as needed.
4. Select
Save .
Содержание KEITHLEY 4200A-SCS-PKC
Страница 1: ...Model 4200A SCS PKC High Power I V C V Parameter Analyzer Quick Start Guide...
Страница 17: ......
Страница 25: ...I V Test...