Tektronix Keithley 2601B-PULSE Скачать руководство пользователя страница 14

Turn the output on 

Turn the output on by pressing the 

OUTPUT ON/

OFF

 control. The ON/OFF indicator LED lights and 

measurements begin.

Observe measurements

Observe the measured voltage on the main area of 

front-panel display. The readings should be very close to the 

20 V source value.

Turn the output off 

When you finish making measurements, turn the output 

 

off by pressing the 

OUTPUT

 

ON/OFF

 control. The output 

indicator LED turns off.

NOTE

These  steps  confirm  basic  functionality  of  your  instrument. 

Turn the instrument power OFF now.

To use the pulser feature, you must have remote 

communications set up. Refer to the 

Model 2601B-PULSE 

User’s Manual

 (document number 2601B-PULSE-900-01) for 

information on remote communications setup.

Содержание Keithley 2601B-PULSE

Страница 1: ...Model 2601B PULSE System SourceMeter Instrument Quick Start Guide...

Страница 2: ...r high transient overvoltages often associated with local AC mains connections Certain Keithley measuring instruments may be connected to mains These instruments will be marked as category II or highe...

Страница 3: ...the associated information very carefully before performing the indicated procedure The CAUTION heading in the user documentation explains hazards that coulddamage the instrument Such damage may inva...

Страница 4: ...maximum at 6 V dc 1 A maximum at 40 V dc Pulser feature output region 4 Region maximums 10 A at 20 V Maximum pulse width 1 8 ms Maximum duty cycle 1 Measurement input electrical ratings Measurement c...

Страница 5: ...the features optimization troubleshooting and programming commands of the instrument Information on accessories Software for the 2601B PULSE is also available for download from the Keithley web page...

Страница 6: ...You receive the 2601B PULSE with these accessories and documents 1 Interlock DB 25 male connector kit hardware 2 Power line cord 3 Safety interlock mating connector 4 Two RJ 45 LAN crossover cables 5...

Страница 7: ...allowed voltage in the system is 300 V dc Cover the device under test DUT to protect the operator from flying debris in the event of a system or DUT failure Make sure any test fixture connected to the...

Страница 8: ...ttom of the instrument Position the instrument so that it is easy to reach any disconnecting devices such as the power cord and the power switch Install the 2601B P INT The 2601B PULSE is shipped with...

Страница 9: ...1B P INT connections firmly onto the terminal strips on the rear panel of the 2601B PULSE 6 Use the two 6 32 in screws provided with the 2601B P INT to secure it to the rear panel of the 2601B PULSE W...

Страница 10: ...vent of a failure not using a properly grounded protective earth and grounded outlet may result in personal injury or death due to electric shock Do not replace detachable mains supply cords with inad...

Страница 11: ...nt is turned on you should see A series of dots All segments of the display light A brief display showing the instrument model 2601B PULSE Line frequency detection and other startup checks The entire...

Страница 12: ...t source function range and level 1 Press the SRC key to display a blinking character in the SrcA value field Confirm that mV is displayed if not press the SRC key again 2 While that character is stil...

Страница 13: ...is 10 0000 mA SrcA 20 0000 V LimA 10 0000mA AUTO Source limit set to 10 0000mA Set measurement function and range 1 Press the MEAS key until the V voltage measurement function is selected In the foll...

Страница 14: ...20 V source value Turn the output off When you finish making measurements turn the output off by pressing the OUTPUT ON OFF control The output indicator LED turns off NOTE These steps confirm basic f...

Страница 15: ...I see an error message when I turn the instrument on If an error message is displayed press the EXIT LOCAL key The 2601B PULSE will return to the default display screen For detailed information about...

Страница 16: ...and foreign patents issued and pending Information in this publication supersedes that in all previously published material Specification and price change privileges reserved TEKTRONIX and TEK are reg...

Отзывы: