Tektronix Keithley 2601B-PULSE Скачать руководство пользователя страница 13

Test

5.  Turn the 

navigation wheel

 to set the source value to 

20.0000 V

,

 then press the 

navigation wheel

 to enter 

the selection and exit EDIT mode.

SrcA:+20.0000 V LimA:1.00000  A

AUTO

Source value set to 20.0000 V

Set the source limit 

1.  Press the 

LIMIT

 key to display a blinking character in the 

LimA

 value field.

2.  Press the down 

RANGE

 key to select the 10 mA limit 

range. Verify that the source limit value in the LimA field 

is 

10.0000 mA

.

SrcA:+20.0000 V LimA:10.0000mA

AUTO

Source limit set to 10.0000mA

Set measurement function and range 

1.  Press the 

MEAS

 key until the V (voltage) measurement 

function is selected. In the following figure, the 

measurement function is set to 

V

.

SrcA:+20.0000 V LimA:10.0000mA

AUTO

Measurement function set to voltage (V)

2.  Press the 

AUTO

 key as many times as needed to select 

the AUTO range function. When AUTO is selected, the 

2601B-PULSE automatically selects the best range for 

the measured value. You briefly see the display shown 

below, then the main display screen reappears.

SrcA:+20.0000 V LimA:10.0000mA

AUTO

AUTO

 

Содержание Keithley 2601B-PULSE

Страница 1: ...Model 2601B PULSE System SourceMeter Instrument Quick Start Guide...

Страница 2: ...r high transient overvoltages often associated with local AC mains connections Certain Keithley measuring instruments may be connected to mains These instruments will be marked as category II or highe...

Страница 3: ...the associated information very carefully before performing the indicated procedure The CAUTION heading in the user documentation explains hazards that coulddamage the instrument Such damage may inva...

Страница 4: ...maximum at 6 V dc 1 A maximum at 40 V dc Pulser feature output region 4 Region maximums 10 A at 20 V Maximum pulse width 1 8 ms Maximum duty cycle 1 Measurement input electrical ratings Measurement c...

Страница 5: ...the features optimization troubleshooting and programming commands of the instrument Information on accessories Software for the 2601B PULSE is also available for download from the Keithley web page...

Страница 6: ...You receive the 2601B PULSE with these accessories and documents 1 Interlock DB 25 male connector kit hardware 2 Power line cord 3 Safety interlock mating connector 4 Two RJ 45 LAN crossover cables 5...

Страница 7: ...allowed voltage in the system is 300 V dc Cover the device under test DUT to protect the operator from flying debris in the event of a system or DUT failure Make sure any test fixture connected to the...

Страница 8: ...ttom of the instrument Position the instrument so that it is easy to reach any disconnecting devices such as the power cord and the power switch Install the 2601B P INT The 2601B PULSE is shipped with...

Страница 9: ...1B P INT connections firmly onto the terminal strips on the rear panel of the 2601B PULSE 6 Use the two 6 32 in screws provided with the 2601B P INT to secure it to the rear panel of the 2601B PULSE W...

Страница 10: ...vent of a failure not using a properly grounded protective earth and grounded outlet may result in personal injury or death due to electric shock Do not replace detachable mains supply cords with inad...

Страница 11: ...nt is turned on you should see A series of dots All segments of the display light A brief display showing the instrument model 2601B PULSE Line frequency detection and other startup checks The entire...

Страница 12: ...t source function range and level 1 Press the SRC key to display a blinking character in the SrcA value field Confirm that mV is displayed if not press the SRC key again 2 While that character is stil...

Страница 13: ...is 10 0000 mA SrcA 20 0000 V LimA 10 0000mA AUTO Source limit set to 10 0000mA Set measurement function and range 1 Press the MEAS key until the V voltage measurement function is selected In the foll...

Страница 14: ...20 V source value Turn the output off When you finish making measurements turn the output off by pressing the OUTPUT ON OFF control The output indicator LED turns off NOTE These steps confirm basic f...

Страница 15: ...I see an error message when I turn the instrument on If an error message is displayed press the EXIT LOCAL key The 2601B PULSE will return to the default display screen For detailed information about...

Страница 16: ...and foreign patents issued and pending Information in this publication supersedes that in all previously published material Specification and price change privileges reserved TEKTRONIX and TEK are reg...

Отзывы: