Model 2231A-30-3 Triple-Channel DC Power Supply Performance Verification
Section 3: Performance Verification
077102700/ July 2014
3-29
3. Power on the DUT and test oscilloscope.
Ensure the warm-up criteria have been met, as described in the Performance Verification Conditions
topic.
4. Set up the equipment as shown.
Figure 5: 20 MHz current noise test setup
5. Set up the oscilloscope as follows:
a. 1 mV/division
b. 1 MΩ input resistance
c. 20 MHz bandwidth limit
d. AC coupled
e. Line trigger
f. 2 ms/div
g. Set to measure V pk-pk and V RMS
6. Set the channel under test channel under test to the 100% full scale output voltage.
7. Set the channel under test to the 100% full scale output current.
8. Turn the DUT output on.
Do not touch the load resistor. The load resistor may become hot enough to cause burns.