Commands Listed in Alphabetical Order
Group
Source
Syntax
[SOURce:]OUTPut:TIMer[:STATe] {0|1|ON|OFF}
[SOURce:]OUTPut:TIMer[:STATe]?
Related Commands
,
Arguments
0
or
OFF
turns the output timer off.
1
or
ON
turns the output timer on.
Returns
0|1
Examples
To activate the timer,
fi
rst send
OUTPUT:TIMER:STATE ON
, then send
OUTPUT:STATE ON
.
To turn the timer off, send
OUTPUT:TIMER:STATE OFF
.
[SOURce:]VOLTage[:LEVel]:DOWN[:IMMediate][:AMPLitude] (No Query Form)
This command is used to decrease the voltage level of the currently selected
channel by a step. The voltage step value can be set by the following command:
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
Group
Source
Syntax
[SOURce:]VOLTage[:LEVel]:DOWN[:IMMediate][:AMPLitude]
Related Commands
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
,
[SOURce:]VOLTage[:LEVel][IMMediate][:AMPLitude]
LEVel]:UP[:IMMediate][:AMPLitude]
[SOURce:]VOLTage[:LEVel][IMMediate][:AMPLitude]
This command is used to set the voltage level of the of the currently selected
channel. The units are V, mV or kV.
Group
Source
2-32
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual
Содержание KEITHLEY 2200 Series
Страница 2: ......
Страница 6: ......
Страница 11: ...Getting Started...
Страница 12: ......
Страница 65: ...Status and Events...
Страница 66: ......
Страница 80: ...Status and Events 3 14 Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual...
Страница 81: ...Appendices...
Страница 82: ......
Страница 83: ...Appendix A ASCII Code Chart Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual A 1...
Страница 84: ...Appendix A ASCII Code Chart A 2 Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual...
Страница 100: ...Appendix C Default Setup C 2 Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual...
Страница 103: ......