Tektronix Keithley 2182A Скачать руководство пользователя страница 7

Unpack

Connect the instrument

Important test system safety information

This product is sold as a stand-alone instrument that may 

become part of a system that could contain hazardous 

voltages and energy sources. It is the responsibility of the 

test system designer, integrator, installer, maintenance 

personnel, and service personnel to make sure the system is 

safe during use and is operating properly.

You must also realize that in many test systems a single 

fault, such as a software error, may output hazardous signal 

levels even when the system indicates that there is no 

hazard present.

It is important that you consider the following factors in your 

system design and use:

• 

The international safety standard IEC 61010-1 defines 

voltages as hazardous if they exceed 30 V

RMS

 and 

42.4 V

PEAK

 or 60 V dc for equipment rated for dry 

locations. Keithley Instruments products are only rated for 

dry locations.

• 

Read and comply with the specifications of all instruments 

in the system. The overall allowed signal levels may be 

constrained by the lowest rated instrument in the system. 

For example, if you are using a 500 V power supply with 

a 300 V dc rated switch, the maximum allowed voltage in 

the system is 300 V dc.

•  Cover the device under test (DUT) to protect the operator 

from flying debris in the event of a system or DUT failure.

• 

Make sure any test fixture connected to the system 

protects the operator from contact with hazardous 

voltages, hot surfaces, and sharp objects. Use 

shields, barriers, insulation, and safety interlocks to 

accomplish this.

•  Double-insulate all electrical connections that an operator 

can touch. Double insulation ensures the operator is 

still protected even if one insulation layer fails. Refer to 

IEC 61010-1 for specific requirements.

•  Make sure all connections are behind a locked cabinet 

door or other barrier. This protects the system operator 

from accidentally removing a connection by hand and 

exposing hazardous voltages. Use high-reliability fail-safe 

interlock switches to disconnect power sources when a 

test fixture cover is opened.

•  Where possible, use automatic handlers so that operators 

are not required to access the DUT or other potentially 

hazardous areas.

•  Provide training to all users of the system so that they 

understand all potential hazards and know how to protect 

themselves from injury.

•  In many systems, during power up, the outputs may be in 

an unknown state until they are properly initialized. Make 

sure the design can tolerate this situation without causing 

operator injury or hardware damage.

Содержание Keithley 2182A

Страница 1: ...Model 2182A Nanovoltmeter Quick Start Guide...

Страница 2: ...high transient overvoltages often associated with local AC mains connections Certain Keithley measuring instruments may be connected to mains These instruments will be marked as category II or higher...

Страница 3: ...read the associated information very carefully before performing the indicated procedure The CAUTION heading in the user documentation explains hazards that could damage the instrument Such damage ma...

Страница 4: ...100 V ac 120 V ac 220 V ac 240 V ac 50 Hz 60 Hz and 400 Hz automatically sensed at power up Power consumption 22 VA Operating environment Specified for 0 C to 50 C specified to 80 relative humidity a...

Страница 5: ...umentation for the 2182A instrument is available for download on the Keithley web page at tek com en products keithley The 2182A documentation includes Quick Start Guide This document It provides unpa...

Страница 6: ...hipping agent immediately 6 There may be a protective film over the display lens which can be removed You receive the 2182A with these accessories and documents Power line cord DeoxIT Contact Cleaner...

Страница 7: ...allowed voltage in the system is 300 V dc Cover the device under test DUT to protect the operator from flying debris in the event of a system or DUT failure Make sure any test fixture connected to the...

Страница 8: ...tch Connect line power The 2182A operates from a line voltage of 100 V ac 120 V ac 220 V ac or 240 V ac at line frequencies of 45 Hz to 66 Hz or 360 Hz to 440 Hz It automatically senses line frequency...

Страница 9: ...instrument by pressing the front panel POWER switch to the on position You must turn on the 2182A and allow it to warm up for at least 2 hours to achieve rated accuracies Power up sequence On power up...

Страница 10: ...t or above 42 VPEAK To prevent electric shock that could result in personal injury or death never make or break connections while hazardous voltage is present CAUTION Exceeding the following limits ma...

Страница 11: ...You can use this cable to make voltage measurements and temperature measurements that use an external simulated reference junction To make voltage connections clamp the cleaned copper lugs of the cabl...

Страница 12: ...in a test circuit The following figure shows typical connections to measure a DUT using a single channel The following figure shows typical connections to make comparison measurements of two devices i...

Страница 13: ...hows connections using the internal reference junction The thermocouple wires must be soldered directly to the CHANNEL connector The following figure shows temperature only connections using an ice ba...

Страница 14: ...tion Press SHIFT then MX B to choose the mx b function Make a voltage measurement The following test verifies basic operation of the 2182A In this test you use the front panel controls to make a volta...

Страница 15: ...e thermocouple type that you are using to measure temperature 6 Select JUNC and select the type of reference measurement To reference measurements to the internal reference junction Select INTRNL To r...

Страница 16: ...sconnected Select DCV1 or DCV2 as appropriate Press REL to null offsets Repeat for other channel if needed Connect the source and make the measurement Keep the 2182A and test circuit away from electri...

Страница 17: ...Next steps For more information refer to the Keithley Instruments website tek com en products keithley You can find support and additional information about the instrument in the Model 2182A Nanovolt...

Страница 18: ...d by U S and foreign patents issued and pending Information in this publication supersedes that in all previously published material Specification and price change privileges reserved TEKTRONIX and TE...

Отзывы: