Voltage measurement connections
You can perform single or dual channel voltage
measurements. The dual channel feature of the 2182A
allows you to make comparison measurements in a
test circuit.
The following figure shows typical connections to measure a
DUT using a single channel.
The following figure shows typical connections to make
comparison measurements of two devices in a test circuit.
For this measurement configuration, there is no voltage
differential between the two measurement channels.
Channel 2 HI is connected directly to Channel 1 LO.