Adjustment Procedures
This section contains an adjustment procedure for your instrument. The purpose
of this procedure is to return the instrument conformance to its speci
fi
cations.
Adjustment interval.
The voltage and timing references inside the instrument
are very stable over time and should not need routine adjustment. The only
time you should perform the
Adjustment Procedures
is if the instrument fails
any of the mainframe performance veri
fi
cation checks provided in the
DSA8300
Speci
fi
cations and Performance Veri
fi
cation
manual.
Adjustment environment.
The instrument must be adjusted in a 20°C to 30 °C
ambient temperature environment. The instrument and signal source must warm
up at least 20 minutes in this environment before you begin the adjustment
procedure.
Required Equipment
The adjustment procedure requires the speci
fi
c test equipment and materials listed
in the following table.
Category
Speci
fi
c equipment required
Quantity
Signal source
Signal generator 50 MHz to 1 GHz,
≤
1 ppm frequency accuracy
1 ea
Frequency counter
Frequency resolution 12 digits
1 ea
Meter
Digital Multimeter, with 6.5 plus digits
1 ea
PC-compatible computer with National Instruments GPIB Controller card and
software, running Windows 95/98 or NT
1 ea
Instrument controller (only
one of these con
fi
gurations
required)
National Instruments USB GPIB Controller card for Windows 98 and National
Instruments NI-488.2 for Windows software (to install in the instrument)
1 ea
Divider
Power, 50
Ω
, SMA "T" male Tektronix 015-0565-00
1 ea
Adapter
SMA "T", male to 2 SMA female Tektronix part number 015-1016-00
1 ea
Adapter
SMA male to BNC female, Tektronix part number 015-0554-00
1 ea
Adapter
BNC to dual banana plug, Tektronix part number 103-0095-00
1 ea
Coaxial cables
50
Ω
, male-to-male SMA connectors
3 ea
GPIB cables
GPIB cable, 1 m min length, Tektronix part number 002-0991-01
1 ea
1
Depending on other USB devices attached to the instrument, a USB hub may also be required. Consult your National
Instruments documentation for more information.
DSA8300 Digital Serial Analyzer and Modules Service Manual
3–1
Содержание DSA8300
Страница 1: ...xx DSA8300 Digital Serial Analyzer and Modules ZZZ Service Manual P077057200 077 0572 00 ...
Страница 2: ......
Страница 3: ...DSA8300 Digital Serial Analyzer and Modules ZZZ Service Manual xx www tektronix com 077 0572 00 ...
Страница 6: ......
Страница 17: ...Operating Information ...
Страница 18: ......
Страница 21: ...Theory of Operation ...
Страница 22: ......
Страница 38: ...Theory of Operation 2 16 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 39: ...Adjustment Procedures ...
Страница 40: ......
Страница 46: ...Adjustment Procedures 3 6 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 47: ...Maintenance ...
Страница 48: ......
Страница 109: ...Replaceable Parts List ...
Страница 110: ......
Страница 115: ...Replaceable Parts List Figure 5 1 External parts DSA8300 Digital Serial Analyzer and Modules Service Manual 5 5 ...
Страница 117: ...Replaceable Parts List Figure 5 2 Drives DSA8300 Digital Serial Analyzer and Modules Service Manual 5 7 ...
Страница 121: ...Replaceable Parts List Figure 5 4 ATX assembly DSA8300 Digital Serial Analyzer and Modules Service Manual 5 11 ...
Страница 123: ...Replaceable Parts List Figure 5 5 Power supply DSA8300 Digital Serial Analyzer and Modules Service Manual 5 13 ...
Страница 125: ...Replaceable Parts List Figure 5 6 Acquisition system DSA8300 Digital Serial Analyzer and Modules Service Manual 5 15 ...
Страница 138: ...Replaceable Parts List Figure 5 17 80A06 module 5 28 DSA8300 Digital Serial Analyzer and Modules Service Manual ...