Clear and Sanitize Procedures

Data Export Devices

The following table lists the data export devices in the standard instrument and
listed modules. Detailed procedures to disable these devices, if any, are shown
following the table.

Table 6: Data Export Devices

Type and
minimum
size

Function

User
modi

fi

able

1

Data input
method

Location

To disable

DSA8200

Removable
hard drive
40 GB

Holds all user-storable data
(waveforms and instrument
settings)

Yes

Application;
Microsoft
Windows

Rear of DSA8200.
Additional drives can
be purchased from
Tektronix.

Remove the hard drive and store it in a
secure area or destroy it.

CD RW /
DVD drive
700 MB

Windows CD RW / DVD
peripheral

Yes

Microsoft
Windows

Front of DSA8200

Remove all CDs. Rewritable CDs can
be formatted, stored in a secure area, or
destroyed. Non-rewritable CDs can either
be stored or destroyed.

USB
host port
(supports
removable
USB

fl

ash

drive)

User storage of reference
waveforms, screen images,
instrument setups, and any other
data through Windows

Yes

User writable
(Windows)

USB host port on
front of instrument
(1 each) and rear of
instrument (4 each).
Files can be deleted
or over-written on
the oscilloscope
or a PC, or USB

fl

ash drive can

be removed and
destroyed.

The USB host port cannot be disabled.

USB
device port

Supports remote control and data
transfer to a PC.

Yes

Not currently
supported

2

USB host port on
front of instrument
(1 each) and rear of
instrument (4 each)

The USB device port is not currently
supported. but may be in the future

LAN
Ethernet
connector

Transfer data

N/A

N/A

Rear panel

N/A

Parallel
port

Windows parallel port

N/A

N/A

Rear panel

N/A

Serial port

Windows serial port

N/A

N/A

Rear panel

N/A

PS/2
Keyboard
and Mouse
Interfaces
(2 each)

Windows serial port

N/A

N/A

Rear panel

N/A

Second
monitor out

Windows second monitor port

N/A

N/A

Rear panel

N/A

Scope out

Scope VGA out

N/A

N/A

Rear panel

N/A

External
audio port

Windows line out audio

N/A

N/A

Rear panel

N/A

GPIB
connector

Transfer data

N/A

N/A

Rear panel

N/A

8

DSA8200 Declassi

fi

cation and Security

Содержание DSA8200 Series

Страница 1: ...xx DSA8200 Digital Serial Analyzer Declassification and Security ZZZ Instructions P071217700 071 2177 00 ...

Страница 2: ......

Страница 3: ...DSA8200 Digital Serial Analyzer Declassification and Security ZZZ Instructions xx www tektronix com 071 2177 00 ...

Страница 4: ...d and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc Contacting Tektronix Tektronix Inc 14150 SW Karl Braun Drive P O Box 500 Beaverton OR 97077 USA For product information sales service and technical support In North America call 1 800 833 ...

Страница 5: ...mory Devices 1 How to Recover from Clearing 80Cxx Module Settings for User Wavelengths and Scales 6 Data Export Devices 8 Built In Security Features 11 Troubleshooting 13 How to Clear or Sanitize a Non Functional Instrument 13 DSA8200 Declassification and Security i ...

Страница 6: ...Table of Contents ii DSA8200 Declassification and Security ...

Страница 7: ...80E10 plug in modules 80A01 80A02 82A04 80A05 and 80A06 plug in modules 80A03 TekConnect Probe Adapter module 80C01 80C02 80C03 80C04 80C05 80C06 80C07 80C08 80C09 80C10 80C11 and 80C12 plug in modules including all options Related Documents DSA8200 Digital Serial Analyzer Service Manual Terms The following terms may be used in this document Clear This removes data on media memory before reusing i...

Страница 8: ...e recovered by other means or technology This is typically used when the device will be moved temporarily or permanently from a secured area to a non secured area Scrub This is equivalent to sanitize User modifiable The user can write to the memory device during normal instrument operation using the instrument interface or remote control Volatile memory Data is lost when the instrument is powered ...

Страница 9: ...Channel SRAM 32 each 24 KB SRAM for Acquisition system channels No Written by processor system Acquisition board Remove power from the instrument for at least 20 seconds SRAM 2 each 352 bytes Channel group microprocessor No Written by processor system Acquisition board Remove power from the instrument for at least 20 seconds 80A03 module Channel SRAM 2 each 1 KB Channel microprocessor No Written b...

Страница 10: ...er Front panel board Not applicable no user data NVRAM 128 kB No Acquisition board Not applicable no user data OTP EPROM 2 each 15 kB Channel group microprocessor No Programmed at factory Acquisition board Not applicable no user data EEPROM 2 each 256 bytes Channel group microprocessor No Not used Acquisition board Not applicable no user data 80E01 80E02 80E03 80E04 80E06 80E07 80E08 80E09 80E10 8...

Страница 11: ...to the upper left large slot Note the upper right large slot may also be used adjusting the channel parameter to CH3 2 Power up the instrument and wait for the oscilloscope application to become active 3 From Table 3 for the applicable module s send each of the programmable interface query commands to the instrument Note that the 80C10 contains two pairs of wavelengths and scales 4 If the user wav...

Страница 12: ...1 TEKPHW CH1 DOUBLE OPTSCALE 2 30 TEKPHW CH1 LONG OPTWAVEL 3 1550 80C12 TEKPHW CH1 DOUBLE OPTSCALE 3 315 Sanitize 80Cxx Modules to Factory Default Settings for User Wavelength s and Scale s The following procedure provides instructions on how to sanitize user wavelength s and scale s to return them to their factory defaults It is recommended that the previous Check 80Cxx Modules User Wavelengths a...

Страница 13: ...ds before performing any other instrument command or operation including powering off the instrument Table 4 Optical Module User Wavelengths Factory Defaults Sanitize Commands Module type all options Programmable interface commands 80C01 TEKPHW CH1 LONG OPTWAVEL 2 1000 TEKPHW CH1 DOUBLE OPTSCALE 2 18 80C02 TEKPHW CH1 LONG OPTWAVEL 2 1000 TEKPHW CH1 DOUBLE OPTSCALE 2 18 80C07B TEKPHW CH1 LONG OPTWA...

Страница 14: ...rface commands to the instrument NOTE The 80C10 contains two pairs of wavelengths and scales 4 After all commands for all applicable modules have been issued to the instrument send the command TEKPHW CH1 EEPROMUPDT to the instrument Wait at least 30 seconds before performing any other instrument command or operation including powering off the instrument 5 Verify that the values match the pre recor...

Страница 15: ...HW CH1 DOUBLE OPTSCALE 1 pre recorded scale value TEKPHW CH1 LONG OPTWAVEL 3 pre recorded wavelength value TEKPHW CH1 DOUBLE OPTSCALE 3 pre recorded scale value 80C11 TEKPHW CH1 LONG OPTWAVEL 2 pre recorded wavelength value TEKPHW CH1 DOUBLE OPTSCALE 2 pre recorded scale value 80C12 TEKPHW CH1 LONG OPTWAVEL 3 pre recorded wavelength value TEKPHW CH1 DOUBLE OPTSCALE 3 pre recorded scale value DSA82...

Страница 16: ...ence waveforms screen images instrument setups and any other data through Windows Yes User writable Windows USB host port on front of instrument 1 each and rear of instrument 4 each Files can be deleted or over written on the oscilloscope or a PC or USB flash drive can be removed and destroyed The USB host port cannot be disabled USB device port Supports remote control and data transfer to a PC Ye...

Страница 17: ...control with attached real time probes No N A Module slots and front of electrical modules TekProbe host ports cannot be disabled Communications will stop if probe is removed 80E01 80E02 80E03 80E04 80E07 80E08 80E09 80E10 modules TekProbe real time probe host interfaces 6 slots 1 or 2 each Communicates setup and control with attached real time probes No N A Front of specific electrical modules 2 ...

Страница 18: ...Clear and Sanitize Procedures 10 DSA8200 Declassification and Security ...

Страница 19: ...Built In Security Features Neither the DSA8200 nor any of the modules covered in this document contain any built in security features DSA8200 Declassification and Security 11 ...

Страница 20: ...Built In Security Features 12 DSA8200 Declassification and Security ...

Страница 21: ...y 80A02 Module This product contains no volatile or non volatile memory 80C01 80C02 80C03 80C04 80C05 80C06 80C07 80C08 80C09 80C10 80C11 80C12 modules including all options The non volatile memory in these products contains no waveform data but does contain a single user storable instrument parameter pair User Wavelength composed of both a wavelength value and an associated scale factor value The...

Отзывы: