Trigger Setup
I
2
C, SPI, CAN, LIN, and FlexRay Bus Trigger Data Matching
Rolling window byte matching for I
2
C, SPI, and FlexRay.
To use a rolling window to trigger on data, you de
fi
ne the
number of bytes to match. Then the oscilloscope uses a rolling window to
fi
nd any match within a packet, with the window rolling
one byte at a time.
For example, if the number of bytes is one, the oscilloscope will attempt to match the
fi
rst byte, second byte, third, and so
on within the packet.
If the number of bytes is two, the oscilloscope will try to match any two consecutive bytes, such as one and two, two and three,
three and four, and so on. If the oscilloscope
fi
nds a match, it will trigger.
With FlexRay, obtain a rolling window match by setting the
Byte Offset
to
Don’t care
. Do this by pressing the front-panel
Trigger
button, followed by setting the lower-bezel
Type
to
Bus
,
Trigger On
to
Data
, pressing
Data
, and entering the byte offset with the
side menu and multipurpose knob
a
..
Speci
fi
c byte matching (non-rolling window matching for a speci
fi
c position in the packet) for I
2
C, SPI, CAN, and
FlexRay.
You can trigger on a speci
fi
c byte for I
2
C, SPI, CAN, and FlexRay in several ways:
For I
2
C and SPI, enter the number of bytes to match the number of bytes in the signal. Then use don’t cares (X) to mask the
bytes that you are not interested in.
For I
2
C, push the lower-bezel
Trigger On
to trigger on
Address/Data
. Push
Address
. On the side-bezel menu, push
Address
,
and rotate multipurpose knobs
a
and
b
as needed. Set the address to don’t cares (X) if you want to mask the address. The
data will be matched starting at the
fi
rst byte without using a rolling window.
For CAN, triggering occurs when the user-selected data input matches the data and quali
fi
er in the signal starting at the
fi
rst
byte. Set the number of bytes to match the number of bytes of interest. Use the data quali
fi
er to perform: =, !=, <, >, >=,
and <= operations. Triggering on identi
fi
er and data always matches the identi
fi
er and data selected by the user, with the
data starting at the
fi
rst byte. No rolling window is used.
154
MSO4000 and DPO4000 Series Oscilloscopes User Manual
Содержание DPO4032
Страница 1: ...x MSO4000 and DPO4000 Series Digital Phosphor Oscilloscopes ZZZ User Manual P071212103 071 2121 03...
Страница 2: ......
Страница 6: ......
Страница 12: ...Table of Contents vi MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Страница 44: ...Installation 18 MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Страница 63: ...Installation 2 Power off your oscilloscope MSO4000 and DPO4000 Series Oscilloscopes User Manual 37...
Страница 67: ...Installation 6 Power on the oscilloscope 7 Push Utility MSO4000 and DPO4000 Series Oscilloscopes User Manual 41...
Страница 294: ...Using Application Modules 268 MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Страница 344: ...Application Examples 318 MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Страница 348: ...Appendix Warranted Specifications 322 MSO4000 and DPO4000 Series Oscilloscopes User Manual...