Installation
Compensating a Passive Voltage Probe
Whenever you attach a passive voltage probe for the
fi
rst time to any input channel, compensate the probe to match it to
the corresponding oscilloscope input channel.
To properly compensate your passive probe:
1.
Follow the steps for the functional
check. (See page 19,
2.
Check the shape of the displayed
waveform to determine if your probe
is properly compensated.
Properly compensated
Under compensated
Over compensated
22